background image

8

 

Caution: Use aseptic techniques whenever the catheter lumen is opened or connected to other devices. 

 

WARNING: Alcohol should not be used to lock, soak or declot polyurethane PICCs because alcohol is known to 

degrade polyurethane catheters over time with repeated and prolonged exposure. 

C. 

Occluded or Partially Occluded Catheter

 

Catheters that present resistance to flushing and aspiration may be partially or completely occluded. Do not flush against 

resistance. If the lumen will neither flush nor aspirate and it has been determined that the catheter is occluded with 

blood, a declotting procedure per institution protocol may be appropriate.

 

17. Central Venous Pressure Monitoring (CVP)

Prior to conducting central venous pressure monitoring:

A. 

Ensure proper positioning of the catheter tip.

•  Flush catheter vigorously with sterile saline.

•  Ensure the pressure transducer is at the level of the right atrium.

•  It is recommended that a continuous infusion of sterile saline (3 mL/hr) is maintained through the catheter while 

measuring CVP to improve accuracy of CVP results.

B. 

Use your institution’s protocols for central venous pressure monitoring procedures.

 

WARNING: CVP Monitoring should always be used in conjunction with other patient assessment metrics when 

evaluating cardiac function.

18. Catheter Removal

A. 

Remove dressing and StatLock® stabilization device or tape securement strips. 

 

Caution: Do not use scissors to remove dressing to minimize the risk of cutting catheter.

B. 

Grasp catheter near insertion site.

C. 

Remove slowly. Do not use excessive force.

D. 

If resistance is felt, stop removal. Apply warm compress and wait 20-30 minutes.

E. 

Resume removal procedure.

F. 

Examine catheter tip to determine that the entire catheter has been removed.

19. Catheters with Sherlock Tip Location System Stylet

Indications for Use: Catheter stylets provide internal reinforcement to aid in catheter placement. The Sherlock™ TLS Stylet 

contains passive magnets that generate a magnetic field. This field can be detected by the Sherlock™ TLS Detector to provide 

the placer rapid feedback on catheter tip location.

Description: The Sherlock™ TLS stylet is made of specially-formulated materials designed to aid in the placement of central 

venous catheters. The stylet material provides internal reinforcement to aid in catheter placement. In addition, the Sherlock™ 

TLS stylet may be used with the detector to provide catheter tip placement information during the insertion procedure.

Note: The Sherlock™ TLS stylet may be used with patients who have cardiac rhythm devices (e.g. pacemakers and defibrillators) 

implanted. When a cardiac rhythm device is present, it is recommended that the Sherlock™ TLS stylet be placed on the 

contralateral side.

Modification of Catheter Length when Using PICC with Sherlock TLS Stylet

Note:  Catheters can be cut to length if a different length is desired due to patient size and desired point of insertion.  Catheter 

depth markings are in centimeters.

A. 

Measure the distance from the insertion site (zero mark) to the desired tip location.

B. 

Loosen the T-lock connector/stylet assembly from the luer hub.

C. 

Withdraw the entire T-lock connector/stylet assembly as one unit. 

D. 

Retract the stylet well behind the catheter cut location. 

E. 

Using a sterile trimming device (e.g. scalpel, scissors, etc.) carefully cut the catheter.   

Caution: The stylet or stiffening wire needs to be well behind the point the catheter is to be cut.  NEVER cut the stylet or 

stiffening wire.

F. 

Inspect cut surface to assure there is no loose material.  

G. 

Re-advance the T-lock connector/stylet assembly locking the connector 

to the luer hub. Assure stylet tip is intact.

H. 

Gently retract the stylet through the locked T-lock connector until the 

stylet tip is contained inside the catheter.

I. 

Prior to insertion, ensure that the stylet tip is contained inside and 

within the catheter but not more than 1 cm from the trimmed end of 

the catheter. [See Figure 1]

 

WARNING: Ensure that the stylet tip does not extend beyond the 

trimmed end of the catheter.  Extension of the stylet tip beyond the 

catheter end, combined with kinking and excessive forces may result in 

vessel damage, stylet damage, difficult removal, stylet tip separation, potential embolism and risk of patient injury. [See 

Figure 2]  

 

Caution: The detector identifies the position of the stylet tip. Ensure that the stylet tip remains inside and within 1 cm 

from the end of the catheter tip. Failure to do so could result in catheter malposition. [See Figure 3]

Figure 1

Correct

1 cm

Incorrect

Figure 2

1 cm

Incorrect

Figure 3

1 cm

Summary of Contents for PowerPICC Provena

Page 1: ...of their respective owners 2018 C R Bard Inc All rights reserved Manufacturer Bard Access Systems Inc 605 North 5600West Salt Lake City UT 84116 USA 801 522 5000 Clinical Information Hotline 1 800 443...

Page 2: ...ue injections Begin appropriate medical intervention immediately Do not wipe the catheter with acetone based solutions tincture of iodine or polyethylene glycol containing ointments These can damage t...

Page 3: ...ed with heparin flush solution As reported in literature anaphylactic or anaphylactic like reactions occur in a small percentage of the population during placement1 positioning1 flushing2 of central v...

Page 4: ...ng the following Air Embolism Bleeding Brachial Plexus Injury Cardiac Arrhythmia Cardiac Tamponade Catheter Erosion Through the Skin Catheter Embolism Catheter Occlusion Catheter Related Sepsis Endoca...

Page 5: ...rd C Insert the safety introducer needle into the desired vein and observe for flashback Alternate Technique A safety IV catheter may be used as an alternate to the safety introducer needle Remove the...

Page 6: ...out advance the catheter into the desired position Caution Never leave stylet or stiffening wire in place after catheter insertion injury may occur Remove stylet or stiffening wire and T lock as appl...

Page 7: ...chor tape with transparent dressing up to hub but not over hub 3 Place 2nd anchor tape sticky side up under hub Wedge tape between hub and wings 4 Chevron anchor tape on top of transparent dressing Du...

Page 8: ...their prevention or treatment Caution If CHG allergy is suspected confirmatory testing is recommended4 5 A Dressing Changes Exit Site Cleaning Caution Do not use scissors to remove dressing to minimi...

Page 9: ...s designed to aid in the placement of central venous catheters The stylet material provides internal reinforcement to aid in catheter placement In addition the Sherlock TLS stylet may be used with the...

Page 10: ...er insertion injury may occur Remove stylet or stiffening wire and T lock as applicable after insertion D For additional information refer to catheter and detector Instructions for Use for the inserti...

Reviews: