background image

4

NOTE: The driver does not contain an on-off switch. The driver automatically recognizes the 
probe type, senses when the probe has been properly inserted and performs an initialization to 
synchronize the driver and probe. The driver sounds one beep and the “PRIME/PIERCE” and 
“SAMPLE” indicator lights display green indicating the probe is properly inserted. During the 
initialization the motor is audible.

6.  After initialization is complete, remove the protective sleeve from the probe.

NOTE: If the driver is left undisturbed for 120 seconds after loading a probe, the driver will 
enter a sleep mode to conserve battery charge. During sleep mode, the “PRIME/PIERCE” and 
“SAMPLE” indicator lights will turn off and the buttons will be deactivated. Once the driver is 
moved, a motion sensor is activated and the driver automatically turns on. The driver will not

 

enter sleep mode if the probe is primed.

Performing a Biopsy

1.  Prepare the biopsy site using appropriate aseptic techniques and adequate local anesthesia. 

Appropriate precautions should be followed for any percutaneous procedure (personal protective 
gear such as gloves and glasses/shields, etc. should be worn). 

2.  Make a small incision or puncture the skin with a scalpel.
3.  Using imaging guidance, insert the probe through the incision and position the tip to the 

appropriate depth:

‡

,IWKHOHVLRQLVJRLQJWREHSLHUFHG3RVLWLRQWKHWLSRIWKHSUREHSUR[LPDOWRWKHHGJHRIWKH

lesion.

‡

,IWKHOHVLRQLVQRWJRLQJWREHSLHUFHG3RVLWLRQWKHVDPSOLQJFKDPEHUWRWKHFHQWHURIWKH

target region. (The center of the sampling chamber is approximately 17mm from the tip of the 
probe.)

4.  If piercing is desired, prior to inserting the probe into the breast press the  

“PRIME/PIERCE”

 button to retract the probe 20-22mm. After inserting the probe press the 

“PRIME/PIERCE” 

button to advance the probe 20-22mm into the region of interest prior to taking 

WKH¿UVWWLVVXHVDPSOH

127(,PDJHJXLGDQFHVKRXOGEHXVHGWRFRQ¿UPSUREHSRVLWLRQUHODWLYHWRWKHWDUJHWUHJLRQWR

be sampled.
NOTE: Do not leave the system unattended with the probe in the primed state.
NOTE: During the initial positioning of the probe, it is recommended to advance the probe into 
the tissue using the optional piercing mode. During subsequent samplings, the physician may 
wish to forego this step.
NOTE: The probe may be primed for piercing while inside the patient if additional piercing shots 
are desired to access dense tissue or lesions.

5.  Begin tissue sampling by pressing the 

“SAMPLE” 

button. The sampling process proceeds 

automatically: A vacuum is created, the outer cannula of the probe is automatically retracted, 
tissue is drawn by vacuum into the sampling chamber and the outer cannula is released cutting 
the tissue. The tissue sample is then transported to the sample container. During this process 
the 

“SAMPLE” 

indicator light continuously illuminates orange and the motor will be audible. The 

sample container illuminates to aid in visualizing the tissue sample. After the tissue sample is 
deposited into the sample container the system resets to acquire another sample if desired.

NOTE: Do not press on the sample container or remove the sample container while acquiring 
tissue samples. If the system detects an error related to the sample container it will illuminate 

WKH6$03/(LQGLFDWRUÀDVKLQJRUDQJHDQGVRXQGEHHSV5HPRYHWKHVDPSOHFRQWDLQHUIURP

the probe, clear any tissue underneath the container or from the probe where the container is 

VHDWHGLQVHUWWKHFRQWDLQHUEDFNLQWRWKHSUREH7KH6$03/(OLJKWZLOOÀDVKJUHHQWRLQGLFDWH

that the sample container is properly inserted. Press the SAMPLE button to resume. The system 
will show the SAMPLE light continuous green. If error is not resolved see Troubleshooting.

6.  To acquire multiple samples, repeat Step 5 above.

NOTE: Collecting multiple tissue samples may help minimize the risk of a false negative biopsy.

 $IWHUREWDLQLQJWKH¿QDOWLVVXHVDPSOHUHPRYHWKHSUREHIURPWKHEUHDVWDQGFDUHIRUWKHLQFLVLRQ

site as appropriate.

NOTE: Prior to removing the probe from the breast, the integrated coaxial cannula may be 
detached and remain in the breast to retain a track to the biopsy site while placing a tissue 
marker (Figure 7).

ENGLISH

Summary of Contents for FINESSE ULTRA

Page 1: ...nstrucciones de uso Biopsiedriver Gebruiksaanwijzing Dispositivo de Colheita de Biopsia Instru es de utiliza o Biopsidriver Brugervejledning Biopsistyrenhet Bruksanvisning Biopsialaite K ytt ohjeet Bi...

Page 2: ...egrated coaxial cannula may be detached after the biopsy and remain in the breast to retain a track to the biopsy site when placing a tissue marker Figure 1 Figure 2 B Indications for Use The FINESSE...

Page 3: ...isolation means Do not position the AC adapter and charging stand such that LW LV GLI FXOW WR UHPRYH WKH DGDSWHU IURP WKH ZDOO RXWOHW LI QHHGHG WR UHPRYH PDLQV SRZHU E Precautions 1 Do not use the FI...

Page 4: ...ly 20 22mm into the breast Using the appropriate imaging guidance the probe s SRVW SLHUFH SRVLWLRQ FDQ EH YHUL HG IWHU WKH SLHUFLQJ VHTXHQFH LV FRPSOHWH WKH PRIME PIERCE and SAMPLE indicator lights co...

Page 5: ...be sampled NOTE Do not leave the system unattended with the probe in the primed state NOTE During the initial positioning of the probe it is recommended to advance the probe into the tissue using the...

Page 6: ...VSHFL F FINESSE ULTRA Biopsy Probe Instructions for Use for additional important information NOTE The probe coaxial cannula and sample container are for single use only After use these products may b...

Page 7: ...G DQG WKHQ URWDWLQJ WKH GULYHU RQWR WKH FKDUJLQJ VWDQG DV VKRZQ LQ LJXUH I WKH EDWWHU LV QRW VXI FLHQWO charged to perform a biopsy the battery indicator light will continuously illuminate red If the...

Page 8: ...stuck in sample chamber or probe chamber is open and probe is stuck in driver If procedure is not yet complete detach coaxial from probe to maintain location in breast Remove probe from the breast I V...

Page 9: ...I 6 OLJKW UHPDLQV DVKLQJ RUDQJH remove the biopsy probe reset the driver and replace with a new probe if additional samples are required S Green Flashing None Sample cup has been reinserted observed a...

Page 10: ...d Wear and tear from normal use or defects resulting from misuse of this product are not covered by this limited warranty TO THE EXTENT ALLOWABLE BY APPLICABLE LAW THIS LIMITED PRODUCT WARRANTY IS IN...