– 15 –
Command:
s_GAIN
Set the gain of an image
Function:
This command compensates the gain value of an image, where “n” represents
the value. The setting can be selected from 20 to 260.
A 100 is set at magnification (Gain 1).
As “s_GAIN” cannot be improved when an image where CCC has been saturated
is multiplied by Gain of 100 or less, because digital data after A/D conversion is
multiplied by a multiplier. In such a case, you should control the IRIS.
In addition, as the multiplication is processed before tone conversion, the output
would not change linearly in the case of no-linear conversion.
Command:
s_GNR n
Set the red gain.
s_GNG n
Set the green gain.
s_GNB n
Set the blue gain.
Function:
These commands compensate the green, red, and blue gain values set in the
unit, where “n” represents a setting value. The setting can be selected from 50 to
150.
A 100 is set at magnification (Gain 1).
The value of this command and the value of “s_GAIN” are respectively
independent. Just like “s_GAIN” digital data after A/D conversion is multiplied.