9
Mode of Operation: Continuous.
Intended use
This display is an AMLCD display designed for viewing medical X-ray
images. This unit should not be used near patients and should be kept
outside of 1.83 m perimeter and 2.29 m vertical.
Power connection
Power requirements: The display must be powered using a medical
approved 12VDC power supply (MDRC-1119) or a 24VDC power supply
(MDRC-2120).
The medical approved DC power supply must be powered by the AC
mains voltage.
Power cords
Utilize a UL-listed detachable power cord, 3-wire, type SJ or equivalent,
18 AWG min., rated 300 V min., provided with a hospital-grade type
plug 5-15P configuration for 120V application, or 6-15P for 240V
application.
Do not overload wall outlets and extension cords as this may result in
fire or electric shock.
Mains lead protection (U.S.: Power cord): Power cords should be routed
so that they are not likely to be walked upon or pinched by items
placed upon or against them, paying particular attention to cords at
plugs and receptacles.
External equipment
External equipment intended for connection to signal input/output or
other connectors, shall comply with relevant UL/ EN/ IEC standard (e.g.
UL/EN/IEC 60950 for IT equipment and UL/EN 60601-1 / IEC 60601
series for medical electrical equipment). In addition, all such
Summary of Contents for MDRC Series
Page 1: ...MDRC Series User manual...
Page 3: ...3...
Page 19: ...19 Installation Product Overview Front View...
Page 23: ...23...
Page 25: ...25 User Controls Jog Dial Controls n o...
Page 37: ...37 Product Dimensions MDRC 1119...
Page 41: ...41 Product Dimensions MDRC 2120...
Page 55: ...55 V1 02 August 2008...