K5903146 /03
Nio Color 5MP
43
Immunity test
IEC 60601 test levels
Compliance level
Electromagnetic
environment
–
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%
Electrical fast transient/
burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
100 kHz repetition
frequency
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
100 kHz repetition
frequency
Mains power quality
should be that of a typical
commercial or hospital
environment
Surge
IEC61000-4-5
Line to line: ± 0.5 kV, ± 1
kV
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
Line to line: ± 0.5 kV, ± 1
kV
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
Mains power quality
should be that of a typical
commercial or hospital
environment
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
0% residual voltage for 0.5
period at 0°, 45°, 90°,
135°, 180°, 225°, 270°
and 315°
0% residual voltage for 1
period at 0°
70% residual voltage for
25 periods at 0°
Voltage interruptions: 0%
residual voltage for 250
periods at 0°
0% residual voltage for 0.5
period at 0°, 45°, 90°,
135°, 180°, 225°, 270°
and 315°
0% residual voltage for 1
period at 0°
70% residual voltage for
25 periods at 0°
Voltage interruptions: 0%
residual voltage for 250
periods at 0°
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the Nio Color 5MP
requires continued
operation during power
mains interruptions, it is
recommended that the Nio
Color 5MP be powered
from an uninterruptible
power supply or a battery
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
30 A/m
Not applicable
6
Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment
Conducted RF
IEC 61000-4-6
3 Vrms (6 Vrms in ISM
bands)
150 kHz to 80 MHz
3 Vrms (6 Vrms in ISM
bands)
-
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
3 V/m
Immunity to RF wireless communications equipment
Test
frequency
(MHz)
Band (MHz)
Service
Modulation
Maximum
power (W)
Distance
(m)
Immunity
test level (V/
m)
385
380
–
390
TETRA 400
Pulse
modulation
18 Hz
1.8
0.3
27
450
430
–
470
GMRS 460,
FRS 460
FM ± 5 kHz
deviation
1 kHz sine
2
0.3
28
710
704
–
787
LTE Band
13, 17
Pulse
modulation
217 Hz
0.2
0.3
9
745
6:
Nio Color 5MP doesn
’
t contain susceptible components to magnetic fields
Summary of Contents for MDNC-6121
Page 1: ...ENABLING BRIGHT OUTCOMES MDNC 6121 User Guide 5 8MP high bright color display Nio Color 5MP...
Page 6: ...K5903146 03 Nio Color 5MP 6 Welcome...
Page 16: ...K5903146 03 Nio Color 5MP 16 Installation...
Page 17: ...17 K5903146 03 Nio Color 5MP Daily operation 3...
Page 21: ...21 K5903146 03 Nio Color 5MP Advanced operation 4...
Page 34: ...K5903146 03 Nio Color 5MP 34 Advanced operation...
Page 35: ...35 K5903146 03 Nio Color 5MP Cleaning your display 5...
Page 37: ...37 K5903146 03 Nio Color 5MP Important information 6...