K5903146 /03
Nio Color 5MP
28
4.15.1 About Embedded QA
About
Embedded QA allows you to run a display calibration or compliance test directly from the display using the
OSD menus described in the next sections. Embedded QA will use the front sensor / I-Guard to measure the
necessary luminance levels for either a calibration or compliance test. Various settings for both actions can be
selected from the display
’
s OSD menu. The last results of both actions can be consulted from the OSD.
Embedded QA or MediCal QAWeb?
Embedded QA is not a replacement for the Barco MediCal QAWeb solution.
Although Embedded QA is a reliable option to perform a simple calibration or compliance test, Barco still
highly recommends MediCal QAWeb as the solution of choice for calibration and QA. Medical QAWeb brings
many benefits such as centralized asset management, the ability to schedule tasks, remote management,
automated reporting, alerting and specific support of regional QA standards such as DIN 6868-57, JESRA and
AAPM TG18. That's why MediCal QAWeb Agent acts as the master for all supported displays from the
moment it is installed and running. MediCal QAWeb Agent will take over from Embedded QA and overwrite
any settings which were applied by Embedded QA.
4.15.2 DICOM status report
About DICOM status report
Following information is available:
DICOM Compliance Status (status since last compliance check):
•
Compliance status:
Shows if the current DICOM curve is compliant or not.
•
Maximum error:
Shows the maximum error of the current DICOM curve. This is the deviation compared to
a perfect DICOM.
•
Error threshold:
Shows the error threshold. This is the maximum error allowed before a DICOM
calibration is required.
•
Time elapsed since latest compliance check:
Shows the backlight runtime since last compliance check.
•
Display Function:
Shows the current display function.
•
Ambient light compensation:
Shows the ambient light compensation status.
•
Reading Room:
Shows the selected reading room.
•
Luminance:
Shows the measured luminance.
•
Black luminance:
Shows the measured black luminance.
DICOM Calibration Status:
•
No calibration executed yet:
No other information is visible
•
Calibration executed:
When the calibration is executed, the following extra information is shown:
Backlight runtime elapsed since latest calibration, display function, ambient light compensation, reading
room.
Current DICOM Settings
•
Display Function:
Shows the current display function.
•
Ambient Light Compensation:
Shows the ambient light compensation status.
•
Reading room:
Shows the selected reading room.
To retrieve the DICOM status report:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Embedded QA
menu.
3.
Select
DICOM status report
to make the information visible on the screen.
Summary of Contents for MDNC-6121
Page 1: ...ENABLING BRIGHT OUTCOMES MDNC 6121 User Guide 5 8MP high bright color display Nio Color 5MP...
Page 6: ...K5903146 03 Nio Color 5MP 6 Welcome...
Page 16: ...K5903146 03 Nio Color 5MP 16 Installation...
Page 17: ...17 K5903146 03 Nio Color 5MP Daily operation 3...
Page 21: ...21 K5903146 03 Nio Color 5MP Advanced operation 4...
Page 34: ...K5903146 03 Nio Color 5MP 34 Advanced operation...
Page 35: ...35 K5903146 03 Nio Color 5MP Cleaning your display 5...
Page 37: ...37 K5903146 03 Nio Color 5MP Important information 6...