5. Important information
Electromagnetic emissions
The Nio Color 2MP 23" is intended for use in the electromagnetic environment speci
fi
ed below. The cus-
tomer or the user of the Nio Color 2MP 23" should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment –
Guidance
RF emissions
CISPR 11
Group 1
The Nio Color 2MP 23" uses
RF energy only for its internal
function. Therefore, its RF
emissions are very low and are
not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class D
Voltage
fl
uctuations/
fl
icker
emissions
IEC 61000-3-3
Complies
The Nio Color 2MP 23" is suitable
for use in all establishments,
including domestic establishments
and those directly connected to
the public low-voltage power
supply network that supplies
buildings used for domestic
purposes.
This Nio Color 2MP 23" complies with appropriate medical EMC standards on emissions to, and interfer-
ence from surrounding equipment. Operation is subject to the following two conditions: (1) this device
may not cause harmful interference, and (2) this device must accept any interference received, including
interference that may cause undesired operation.
Interference can be determined by turning the equipment off and on.
If this equipment does cause harmful interference to, or suffer from harmful interference of, surrounding
equipment, the user is encouraged to try to correct the interference by one or more of the following mea-
sures:
•
Reorient or relocate the receiving antenna or equipment.
•
Increase the separation between the equipment and receiver.
•
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
•
Consult the dealer or an experienced technician for help.
Electromagnetic immunity
The Nio Color 2MP 23" is intended for use in the electromagnetic environment speci
fi
ed below. The cus-
tomer or the user of the Nio Color 2MP 23" should assure that it is used in such an environment.
Immunity test
IEC 60601
Test levels
Compliance level
Electromagnetic
environment –
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
Floors should be wood,
concrete or ceramic tile.
If
fl
oors are covered with
synthetic material, the
relative humidity should
be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
± 2kV for power supply
lines
± 1kV for input/ output
lines
± 2kV for power supply
lines
± 1kV for input/ output
lines
Mains power quality
should be that of a typical
commercial or hospital
environment
Surge
IEC61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality
should be that of a typical
commercial or hospital
environment
26
K5902134 NIO COLOR 2MP 23" 05/07/2017