6. Important information
Power connection – Equipment with external 24 VDC power supply
•
Power requirements: The equipment must be powered using the delivered medical approved 24 VDC
(
) SELV power supply.
•
The medical approved DC (
) power supply must be powered by the AC mains voltage.
•
The power supply is speci
fi
ed as a part of the ME equipment or combination is speci
fi
ed as a ME
system.
•
To avoid the risk of electric shock, this equipment must only be connected to a supply mains with
protective earth.
•
The equipment should be installed near an easily accessible outlet.
•
The equipment is intended for continuous operation.
Power cords:
•
Europe: H05VV-F or H05VVH2-F PVC cord with appropriate EU plug.
US and Canada: “hospital grade” cord-set has to be used, provided with instructions to indicate that
grounding reliability can be achieved only when the equipment is connected to an equivalent recep-
tacle marked hospital only or hospital grade. These instructions need to be marked either on the
equipment or on a tag on the power cord.
•
Do not overload wall outlets and extension cords as this may result in
fi
re or electric shock.
•
Mains lead protection (U.S.: Power cord): Power cords should be routed so that they are not likely to
be walked upon or pinched by items placed upon or against them, paying particular attention to cords
at plugs and receptacles.
•
Use a power cord that matches the voltage of the power outlet, which has been approved and complies
with the safety standard of your particular country.
Transient over-voltage
If the device is not used for a long time, disconnect it from the AC inlet to avoid damage by transient
over-voltage.
To fully disengage the power to the device, please disconnect the power cord from the AC inlet.
Connections
Any external connection with other peripherals must follow the requirements of clause 16 of IEC60601-1
3rd ed. or Table BBB.201 of IEC 60601-1-1 for the medical electrical systems.
Water and moisture
The equipment is IP21 compliant. The monitor front side only is IPx5 compliant.
The external power supply is approved for IPx0. The power supply must be mounted in
a
fl
at position for best resistance to
fl
uids.
Ventilation
Do not cover or block any ventilation openings in the cover of the set. When installing the device in a
cupboard or another enclosed location, heed the necessary space between the set and the sides of the
cupboard.
36
K5902113
(
451920611383
)
E192HSA
27/11/2017
Summary of Contents for E192HSA
Page 1: ...E192HSA User Guide 19 inch medical grade color display K5902113 451920611383 01 27 11 2017...
Page 12: ...2 Parts controls and connectors 10 K5902113 451920611383 E192HSA 27 11 2017...
Page 16: ...3 Display installation 14 K5902113 451920611383 E192HSA 27 11 2017...
Page 22: ...4 Daily operation 20 K5902113 451920611383 E192HSA 27 11 2017...
Page 36: ...5 Advanced operation 34 K5902113 451920611383 E192HSA 27 11 2017...