Emissions test
Compliance
Electromagnetic environment
–
Guidance
supplies buildings used for
domestic purposes.
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
Complies
This AMM240ED complies with appropriate medical EMC standards on emissions to, and interference from
surrounding equipment. Operation is subject to the following two conditions: (1) this device may not cause
harmful interference, and (2) this device must accept any interference received, including interference that
may cause undesired operation.
Interference can be determined by turning the equipment off and on.
If this equipment does cause harmful interference to, or suffer from harmful interference of, surrounding
equipment, the user is encouraged to try to correct the interference by one or more of the following measures:
•
Reorient or relocate the receiving antenna or equipment.
•
Increase the separation between the equipment and receiver.
•
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
•
Consult the dealer or an experienced technician for help.
Electromagnetic immunity
The AMM240ED is intended for use in the electromagnetic environment specified below. The customer or the
user of the AMM240ED should assure that it is used in such an environment.
Immunity test
IEC 60601 test levels
Compliance level
Electromagnetic
environment
–
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%
Electrical fast transient/
burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
100 kHz repetition
frequency
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
100 kHz repetition
frequency
Mains power quality
should be that of a typical
commercial or hospital
environment
Surge
IEC61000-4-5
Line to line: ± 0.5 kV, ± 1
kV
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
Line to line: ± 0.5 kV, ± 1
kV
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
Mains power quality
should be that of a typical
commercial or hospital
environment
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
0% residual voltage for 1
period at 0°
70% residual voltage for
25/30 cycles periods at 0
⁰
Voltage interruptions: 0%
residual voltage for 250/
300 cycles periods at 0
⁰
0% residual voltage for 1
period at 0°
70% residual voltage for
25/30 cycles periods at 0
⁰
Voltage interruptions: 0%
residual voltage for 250/
300 cycles periods at 0
⁰
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the AMM240ED requires
continued operation
during power mains
interruptions, it is
recommended that the
AMM240ED be powered
from an uninterruptible
power supply or a battery
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
30 A/m
Not applicable
13
Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
13. AMM240ED doesn
’
t contain susceptible components to magnetic fields
Summary of Contents for AMM240ED
Page 1: ...ENABLING BRIGHT OUTCOMES User Guide AMM240ED...
Page 11: ...11 R5912858 05 AMM240ED Daily operation 3...
Page 14: ...R5912858 05 AMM240ED 14 Daily operation...
Page 15: ...15 R5912858 05 AMM240ED Advanced operation 4...
Page 21: ...21 R5912858 05 AMM240ED Cleaning your display 5...
Page 23: ...23 R5912858 05 AMM240ED Troubleshooting 6...
Page 25: ...25 R5912858 05 AMM240ED Important information 7...
Page 43: ......