Baoding Longer Precision Pump LEAD-2 Longer Operating Manual Download Page 8

Flow Rate Calibration:

Start/Stop Key

Start/Stop Key

Start/Stop

Key

Adjusting Knob

Adjusting

Knob

Adjusting Knob

Adjusting Knob

Adjusting Knob

1. For the first time flow rate calibration, “System Setting” must

be finished. Please see page 6 for setting method.

2. Press the

to begin the testing. Press the

again to stop the testing. The pump stops

automatically and enter “prompt interface” as soon as the
setting time is up. In “prompt interface” press the

to repeat the testing. Press the

to enter

actual volume modification interface. Turn the

to input the actual volume and then press the Adjusting

Knob to return to “1. Calibration” interface.

3. Turn the

to enter “4. Exit” interface. Press the

to return to “Running Interface”. The flow

rate displayed in the “Running Interface” is actual flow rate.
Turn the

to input the required flow rate.

Re-calibrate

Adjust Vol

ID 0.5

INT

123.4mL/min

System Setting

The submenus of “System Setting” menu are “a. Tubing ID”, “b.
Test time”, “c. Remote Control”, “d. Pump I.D. #”, “e. Footswitch”
and “f. Back”.

Calibration Example
Assume the required flow rate is 100 μL/min.

ID 0.5

INT

100.8 L/min

μ

ID 0.5

INT

120.90 L/min

μ

ID 0.5

INT

100 L/min

μ

a. Tubing ID

In “System Setting” interface, press the

to enter

“a. Tubing ID” interface. Press the

to enter

“Select Tubing ID” interface. Turn the

to select

the required tubing inner diameter. Press the

to

save and return back to “a. Tubing ID” interface.

Adjusting knob

Adjusting Knob

Adjusting Knob

Adjusting Knob

Select Tubing ID

1

6.4

3.

4.8

ID 0.5

INT

123.4mL/min

ID 0.5

INT

123.4mL/min

5

6

LEAD - 2 OPERATING MANUAL

LEAD - 2 OPERATING MANUAL

1.Calibration

1.Calibration

1.Calibration

2. ystem Setting

S

2. ystem Setting

S

4. xit

E

4. xit

E

Act Vol:

1234 μL

Press

to exit

Test time

0.5min

Cal Vol:

1234 μL

Press

to begin

Re-calibrate

Adjust Vol

Act Vol:

1234 μL

Press

to exit

1.Calibration

a.Tubing ID

0.5mm

b.Test time

0.5min

c.Remote

Control:on

d. ump I.D.#.

P

1

e.Footswitch

Gated

:

f. ack

B

a.Tubing ID

0.5mm

f. ack

B

Running interface

Running interface

Press
the Knob

Press
the Knob

Press
the Knob

Turn the
Knob

Turn the knob to input
the required flow rate

Turn the knob to input
the actual volume

Running interface

Running interface

Press
the Knob

Press
the Knob

Press
the Knob

Press
the Knob

Press
the Knob

Turn the
Knob

Display the
next screen
automatically
after 2 seconds

Turn the
Knob

To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method

System Setting” must be finished
b e f o r e

c a l i b ra t i o n .

P l e a s e

s e e

page 6 for setting method

Press

Key

/

Display Prompt Interface
after testing

Turn the knob to input the
actual volume

Turn the knob to input
the required flow rate

Press
the Knob

Press
the Knob

Press
the Knob

Press
the Knob

Press the Knob

Turn the
Knob

Turn the
Knob

Turn the
Knob

Turn the
Knob

Turn the
Knob

Turn the
Knob

Turn the
Knob

Running interface

To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method

Running interface

To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method

Press the knob
and then turn
the knob

Turn the knob to select

Summary of Contents for LEAD-2 Longer

Page 1: ...g Longer Precision Pump Co Ltd ADD Building A Chuangye Center Baoding National High Tech Industrial Development Zone Baoding Hebei China 071051 TEL 86 312 3110087 3138553 FAX 86 312 3168553 E mail longer longerpump com Www longerpump com LEAD 2 PERISTALTIC PUMP OPERATING MANUAL ...

Page 2: ...wn the power supply and pull out the plug when meet below circumstances Hold the plug not the power line when pulling out The fluids splash on the body of the pump You think the pump needs to be maintained or repaired Please shut down the power supply before install the external control equipments LEAD 2 OPERATING MANUAL Introduction Operation Panel Basic Operation Running Interface Menu Frame Wor...

Page 3: ...wn the power supply and pull out the plug when meet below circumstances Hold the plug not the power line when pulling out The fluids splash on the body of the pump You think the pump needs to be maintained or repaired Please shut down the power supply before install the external control equipments LEAD 2 OPERATING MANUAL Introduction Operation Panel Basic Operation Running Interface Menu Frame Wor...

Page 4: ... Control Interface P3 Power Socket Reference For detail information about external control function please see page 13 Basic Operation Start Stop Key Press the to start or stop the pump Direction Key Press the to change the rotation direction of the pump This key is valid when the pump is running Adjusting Knob Function 1 Adjust the flow rates or speed in running state Function 2 In menus selectio...

Page 5: ... Control Interface P3 Power Socket Reference For detail information about external control function please see page 13 Basic Operation Start Stop Key Press the to start or stop the pump Direction Key Press the to change the rotation direction of the pump This key is valid when the pump is running Adjusting Knob Function 1 Adjust the flow rates or speed in running state Function 2 In menus selectio...

Page 6: ...setting interface Note In speed Display the calibration function is forbidden In Running Interface press the it will display Adjusting Knob LEAD 2 OPERATING MANUAL LEAD 2 OPERATING MANUAL 1 Calibration 2 ystem Setting S 3 Control mode 4 xit E 1 Calibration a Running direction b Current flow rate and speed c Tubing inner diameter ID0 5 means the inner diameter of the tubing is 0 5 mm d Start Stop s...

Page 7: ...setting interface Note In speed Display the calibration function is forbidden In Running Interface press the it will display Adjusting Knob LEAD 2 OPERATING MANUAL LEAD 2 OPERATING MANUAL 1 Calibration 2 ystem Setting S 3 Control mode 4 xit E 1 Calibration a Running direction b Current flow rate and speed c Tubing inner diameter ID0 5 means the inner diameter of the tubing is 0 5 mm d Start Stop s...

Page 8: ... Knob Adjusting Knob Select Tubing ID 1 6 4 3 4 8 ID 0 5 INT 123 4mL min ID 0 5 INT 123 4mL min 5 6 LEAD 2 OPERATING MANUAL LEAD 2 OPERATING MANUAL 1 Calibration 1 Calibration 1 Calibration 2 ystem Setting S 2 ystem Setting S 4 xit E 4 xit E Act Vol 1234 μL Press to exit Test time 0 5min Cal Vol 1234 μL Press to begin Re calibrate Adjust Vol Act Vol 1234 μL Press to exit 1 Calibration a Tubing ID ...

Page 9: ... Knob Adjusting Knob Select Tubing ID 1 6 4 3 4 8 ID 0 5 INT 123 4mL min ID 0 5 INT 123 4mL min 5 6 LEAD 2 OPERATING MANUAL LEAD 2 OPERATING MANUAL 1 Calibration 1 Calibration 1 Calibration 2 ystem Setting S 2 ystem Setting S 4 xit E 4 xit E Act Vol 1234 μL Press to exit Test time 0 5min Cal Vol 1234 μL Press to begin Re calibrate Adjust Vol Act Vol 1234 μL Press to exit 1 Calibration a Tubing ID ...

Page 10: ...tem Setting interface press and then turn the to enter e Footswitch interface Press the to enter footswitch setting interface Turn the to select footswitch working mode Press the to save and return back to e Footswitch interface Trigger Press footswitch the pump starts running Press footswitch again the pump stops Gated The pump runs as long as the footswitch is pressed Adjusting Knob Adjusting Kn...

Page 11: ...tem Setting interface press and then turn the to enter e Footswitch interface Press the to enter footswitch setting interface Turn the to select footswitch working mode Press the to save and return back to e Footswitch interface Trigger Press footswitch the pump starts running Press footswitch again the pump stops Gated The pump runs as long as the footswitch is pressed Adjusting Knob Adjusting Kn...

Page 12: ...ly When the pump connects the footswitch the on operating panel is invalid no matter the remote control is on or off Start Stop Key ID 0 5 INT 123 4mL min Acceptable Pump Heads and Tubing Pump Head Tubing mm 16 25 17 LEAD15 48 LEAD15 88 LEAD15 24 LEAD15 44 2 4 Inner Diameter 1 0 Wall Thickness 16 25 17 9 10 LEAD 2 OPERATING MANUAL LEAD 2 OPERATING MANUAL 1 Calibration 2 ystem Setting S 1 Calibrati...

Page 13: ...ly When the pump connects the footswitch the on operating panel is invalid no matter the remote control is on or off Start Stop Key ID 0 5 INT 123 4mL min Acceptable Pump Heads and Tubing Pump Head Tubing mm 16 25 17 LEAD15 48 LEAD15 88 LEAD15 24 LEAD15 44 2 4 Inner Diameter 1 0 Wall Thickness 16 25 17 9 10 LEAD 2 OPERATING MANUAL LEAD 2 OPERATING MANUAL 1 Calibration 2 ystem Setting S 1 Calibrati...

Page 14: ...blow to the other side of the tubing If there is not air bubble in water the screws doesn t need to be adjusted If there is air bubble in water adjust the screws until the location where there is not air bubble in water 11 12 LEAD 2 OPERATING MANUAL LEAD 2 OPERATING MANUAL Pump Head Mounting The pump head is mounted on drive before the pump leaves the factory Follow below procedure for changing an...

Page 15: ...blow to the other side of the tubing If there is not air bubble in water the screws doesn t need to be adjusted If there is air bubble in water adjust the screws until the location where there is not air bubble in water 11 12 LEAD 2 OPERATING MANUAL LEAD 2 OPERATING MANUAL Pump Head Mounting The pump head is mounted on drive before the pump leaves the factory Follow below procedure for changing an...

Page 16: ...s below Please contact Longer Company for communication protocol Terminal Definition of Standard External Control Module Analog speed control input for external speed control in external control modules of 4 to 20 mA 0 to 5V and 0 to 10 V External start stop control input When connected to COM the pump runs when connected to high level 5 to 12 V the pump stops External cw ccw control input When co...

Page 17: ...s below Please contact Longer Company for communication protocol Terminal Definition of Standard External Control Module Analog speed control input for external speed control in external control modules of 4 to 20 mA 0 to 5V and 0 to 10 V External start stop control input When connected to COM the pump runs when connected to high level 5 to 12 V the pump stops External cw ccw control input When co...

Page 18: ... If repair or adjustment is necessary within the warranty period the problem will be corrected at no charge if it is not due to misuse or abuse on your part as determined by the manufacturer Repair costs outside the warranty period or those resulting from product misuse or abuse may be invoiced to you Technical Specifications Functions Applicable Pump Head LEAD15 24 LEAD15 44 LEAD15 48 LEAD15 88 F...

Page 19: ... If repair or adjustment is necessary within the warranty period the problem will be corrected at no charge if it is not due to misuse or abuse on your part as determined by the manufacturer Repair costs outside the warranty period or those resulting from product misuse or abuse may be invoiced to you Technical Specifications Functions Applicable Pump Head LEAD15 24 LEAD15 44 LEAD15 48 LEAD15 88 F...

Page 20: ...g Longer Precision Pump Co Ltd ADD Building A Chuangye Center Baoding National High Tech Industrial Development Zone Baoding Hebei China 071051 TEL 86 312 3110087 3138553 FAX 86 312 3168553 E mail longer longerpump com Www longerpump com LEAD 2 PERISTALTIC PUMP OPERATING MANUAL ...

Reviews: