Flow Rate Calibration:
Start/Stop Key
Start/Stop Key
Start/Stop
Key
Adjusting Knob
Adjusting
Knob
Adjusting Knob
Adjusting Knob
Adjusting Knob
1. For the first time flow rate calibration, “System Setting” must
be finished. Please see page 6 for setting method.
2. Press the
to begin the testing. Press the
again to stop the testing. The pump stops
automatically and enter “prompt interface” as soon as the
setting time is up. In “prompt interface” press the
to repeat the testing. Press the
to enter
actual volume modification interface. Turn the
to input the actual volume and then press the Adjusting
Knob to return to “1. Calibration” interface.
3. Turn the
to enter “4. Exit” interface. Press the
to return to “Running Interface”. The flow
rate displayed in the “Running Interface” is actual flow rate.
Turn the
to input the required flow rate.
Re-calibrate
Adjust Vol
↑
ID 0.5
INT
■
123.4mL/min
System Setting
The submenus of “System Setting” menu are “a. Tubing ID”, “b.
Test time”, “c. Remote Control”, “d. Pump I.D. #”, “e. Footswitch”
and “f. Back”.
Calibration Example
Assume the required flow rate is 100 μL/min.
ID 0.5
INT
■
100.8 L/min
μ
ID 0.5
INT
■
120.90 L/min
μ
ID 0.5
INT
■
100 L/min
μ
a. Tubing ID
In “System Setting” interface, press the
to enter
“a. Tubing ID” interface. Press the
to enter
“Select Tubing ID” interface. Turn the
to select
the required tubing inner diameter. Press the
to
save and return back to “a. Tubing ID” interface.
Adjusting knob
Adjusting Knob
Adjusting Knob
Adjusting Knob
Select Tubing ID
1
6.4
3.
4.8
↑
ID 0.5
INT
■
123.4mL/min
ID 0.5
INT
■
123.4mL/min
5
6
LEAD - 2 OPERATING MANUAL
LEAD - 2 OPERATING MANUAL
1.Calibration
1.Calibration
1.Calibration
2. ystem Setting
S
2. ystem Setting
S
4. xit
E
4. xit
E
Act Vol:
1234 μL
Press
to exit
Test time
0.5min
:
Cal Vol:
1234 μL
Press
to begin
Re-calibrate
Adjust Vol
Act Vol:
1234 μL
Press
to exit
1.Calibration
a.Tubing ID
0.5mm
:
b.Test time
0.5min
:
c.Remote
Control:on
d. ump I.D.#.
P
1
e.Footswitch
Gated
:
f. ack
B
a.Tubing ID
0.5mm
:
f. ack
B
Running interface
Running interface
Press
the Knob
Press
the Knob
Press
the Knob
Turn the
Knob
Turn the knob to input
the required flow rate
Turn the knob to input
the actual volume
Running interface
Running interface
Press
the Knob
Press
the Knob
Press
the Knob
Press
the Knob
Press
the Knob
Turn the
Knob
Display the
next screen
automatically
after 2 seconds
Turn the
Knob
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
System Setting” must be finished
b e f o r e
c a l i b ra t i o n .
P l e a s e
s e e
page 6 for setting method
Press
Key
/
■
Display Prompt Interface
after testing
Turn the knob to input the
actual volume
Turn the knob to input
the required flow rate
Press
the Knob
Press
the Knob
Press
the Knob
Press
the Knob
Press the Knob
Turn the
Knob
Turn the
Knob
Turn the
Knob
Turn the
Knob
Turn the
Knob
Turn the
Knob
Turn the
Knob
Running interface
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
Running interface
To switch to “Speed Display”
please see page 9 “Control
Mode” for setting method
Press the knob
and then turn
the knob
Turn the knob to select