POL-200 Semiautomatic Polarimeter 

 

Storing and Recalling Data from Memory 

POL-200 semiautomatic polarimeter allows up to 100 data sets to be stored and recalled.   
 
MEMORY INPUT: 
1.1  During the measurement, click the Save icon, the screen shows a dialog box and wait for entering the sample name. 
 
 

 

 
1.2  The default sample name is SAMPLE. If you need to modify this name, press 

 key, then press the letter keys to enter new name. Press ENTER key   

   

to confirm, the measured value will be saved into the memory. 

 

 

 
 
 

 

Summary of Contents for POL-200

Page 1: ...POL 200 Semiautomatic Polarimeter Instruction Manual BANTE INSTRUMENTS LIMITED...

Page 2: ...ete If any are damaged or missing please contact nearest distributor Test tube 100 and 200 mm Sealing rings Power cable Power On Off Plug in the power supply and press ON OFF key to turn on the polari...

Page 3: ...0 to 220 0mm Temperature Compensation Enable Default Disable Temperature compensation allows the instrument to run at ambient temperature but produces readings as if the system was running at 20 C Du...

Page 4: ...ior to Use Rotate the focus knob on front panel until visual field became legible Zero Correction 1 Take a clean empty tube and place in the sample chamber 2 If the visual field appears the difference...

Page 5: ...ke out the tube from the sample chamber Unscrew the cap remove the sealing ring and glass disc 2 Pour sample solution into the tube screw on the tube cap Wipe away the drops of water on glass disc 3 I...

Page 6: ...red and recalled MEMORY INPUT 1 1 During the measurement click the Save icon the screen shows a dialog box and wait for entering the sample name 1 2 The default sample name is SAMPLE If you need to mo...

Page 7: ...ring the measurement mode the screen shows the stored measured values 2 2 Press Page key the screen will toggle to next page 2 3 If you need to delete stored data press Delete key all of the data will...

Page 8: ...p icon during the measurement mode the screen shows the setup menu 2 Click the Calibration icon the instrument enters the calibration mode 3 Follow the displayed guides to select the number of calibra...

Page 9: ...ows Calibration is completed The instrument will return to measurement mode automatically Calibration is completed Restore Factory Settings If the instrument goes wrong during the use you are able to...

Page 10: ...90 130 Z Accuracy 0 02 0 05 Z Resolution 0 005 Wavelength 589nm Light Source LED Interference filter Test Tube 50 100 200mm Temperature Range 0 100 C Temperature Accuracy 0 5 C Memory Stores up to 10...

Page 11: ...w to speed up these efforts Warranty The warranty period for polarimeter is two years from the date of shipment Above warranty does not cover light source Out of warranty products will be repaired on...