Measurement 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Measuring the High Turbidity Sample 

For the high turbidity samples (>

 

1100 NTU), the solution must be diluted 

before measurement. The dilution water can be obtained by filtering 
distilled water through a < 0.45 µm filter membrane. 
2.1  Repeat steps 1.1 through 1.4 above and record the reading. 
2.2  Calculate the true turbidity of the original sample using the   
 

following formula. 

 
 

T = [T

d

 (V

s

 + V

d

)] / V

s

 

 
 

Where:   

 

T

   

= Ture turbidity of the original sample 

 

T

d

 

= Measured value 

 

 

V

s

   

= Volume of the original sample (ml) 

 

V

d

 = Volume of the dilution water (ml) 

 

 

 

Total Suspended Solids   
Measurement 

The TB200 turbidity meter contains a Total Suspended Solids mode 
and needs to enter a conversion factor before measurement. 
 

Calculating the TSS Factor 

1. 

Rinse the filter disk with distilled water to remove any solids that 

 

may remain. Put the filter disk in a watch glass and dry them in 

 

a 104°C (219°F) drying oven for 1 hour. 

2. 

Remove the filter disk and watch glass and put in a desiccator. 

 

Immediately cover the desiccator. Wait for them decrease to   

 

room temperature. 

3. 

Weigh the filter disk and watch glass and record the mg value 

 

as B. 

4. 

Filter the 100 ml of sample through pre-weighed filter disk. Put 

 

the filter disk and watch glass in a 104°C (219°F) drying oven for 

 

1 hour.   

5. 

Remove the filter disk and watch glass and put in a desiccator. 

 

Immediately cover the desiccator. Wait for them decrease to   

 

room temperature. 

6. 

Weigh the filter disk and watch glass and record the mg value 

 

as A. 

7. 

Calculate the TSS value using the formula below. 

 
 

TSS (mg/L) = (A - B) / 0.1 

 
8. 

Fill the vial with the sample.   

9. 

Repeat turbidity measurement steps

 

1.1 through 1.5 and record 

 

the NTU value. Calculate the TSS conversion factor using the   

 

formula below. 

 
 

Factor = NTU / TSS (mg/L) 

 

Setting the TSS Factor 

1. 

Press and hold the

 

 

key to enter the setup menu. 

2. 

Press the 

 key to select the 

TSS Factor

, press the 

Enter

 key to 

 

confirm. 

3. 

Press the 

 / 

 key to set the conversion factor, press the 

Enter

 

 

key to save. 

 

 

 

 

 

 

 

Selecting the TSS Measurement Mode 

1. 

Press and hold the

 

 

key to enter the setup menu. 

2. 

Press  the 

  key  to select  the 

Measurement  Mode

, press  the 

 

Enter

 key to confirm.   

3. 

Press the 

 / 

 key to select the 

mg/L (TSS)

, press the 

Enter

 

 

key to enter the Total Suspended Solids measurement mode. 

 

 

 

 

 

 

Summary of Contents for TB200

Page 1: ...TB200 Laboratory Turbidity Meter USER MANUAL Bante Instruments Inc ...

Page 2: ...zin standard solutions 0 02 200 500 1000 NTU Two sample vials Lint free cloth 12V DC power adapter Meter Overview 1 Light shield lid 2 Measurement chamber 3 Display 4 Membrane keypad Connectors 1 Socket for power adapter 2 USB B interface to computer Keypad Key Function Switch the meter on or off Exit the calibration settings data logs and return to the measurement mode Start calibration Press and...

Page 3: ...calibration Enable Disable default Reset Reset the meter to factory default settings Note the meter must be recalibrated Enable Disable default Setting a Default Option 1 1 Press and hold the key to enter the setup menu 1 2 Press the key to select a menu item or an option 1 3 Press the Enter key to save To exit the setting without saving changes press the ESC key Setting the Password The password ...

Page 4: ...ow on the meter Always close the light shield lid to prevent the measurement errors from ambient light Wash the sample vial with distilled water after measurement Any residues can cause inaccurate readings Turbidity Calibration The TB200 turbidity meter allows turbidity calibration up to 7 points with minimum of 2 points the default calibration points include the 0 02 10 200 500 1000 1500 and 2000...

Page 5: ...n page 3 Measuring the Low Turbidity Sample For the low turbidity samples 200 NTU we recommend that using the same vial to perform the calibration and measurement 1 1 Rinsethe vial withapproximately10 mlof sample Cap andupend the vial several times Discard the used sample and repeat the rinsing procedure two more times 1 2 Fill the vial with the sample Cap the vial and wipe with a lint free cloth ...

Page 6: ...hem decrease to room temperature 3 Weigh the filter disk and watch glass and record the mg value as B 4 Filter the 100 ml of sample through pre weighed filter disk Put the filter disk and watch glass in a 104 C 219 F drying oven for 1 hour 5 Remove the filter disk and watch glass and put in a desiccator Immediately cover the desiccator Wait for them decrease to room temperature 6 Weigh the filter ...

Page 7: ...g the Data 1 Connect the USB cable to the meter and computer 2 Click the DAS icon the system automatically scans an available communication port and shows a message box Found a port on your computer 3 Click the OK the application starts 4 Click the Connect the screen shows Port is connected 5 Click the OK thenclick the Receive the storeddata will transfer to computer automatically If your computer...

Page 8: ...100 ml volumetric flask and dilute to the mark 200 NTU Mix 10 ml of 4000 NTU standard in a 100 ml volumetric flask and dilute to the mark Mix 50ml of above standard in a 100 ml volumetric flask and dilute to the mark 10 NTU Mix 2 5ml of 200 NTU standard in a 100ml volumetric flask and dilute to the mark Or Mix 2ml of 500 NTU standard in a 100ml volumetric flask and dilute to the mark Optional Acce...

Page 9: ...ter is one year from the date of shipment Above warranty does not cover the turbidity standards and glass vials Out of warranty products will be repaired on a charged basis The warranty on your meter shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the environment specifications of the products For mor...

Reviews: