3   

I   

Calibration 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Factory Reset

 

The Reset feature is used to restore the meter back to factory default 
settings. If enabled, all of the calibration data and selected options   
will be deleted or reset, the meter must be recalibrated. 
 
 

Note, the data logs will not be deleted. 

 
3.1  Press and hold the

 

 

key to enter the setup menu. 

3.2  Press the 

 key to select the 

Reset

, press the 

Enter

 key to   

 

confirm. 

3.3  Press the 

 key to select the 

Restore Factory Settings

. Press the 

 

Enter

 key, the screen shows "Are you sure you want to restore 

 

factory settings?

 

". 

3.4  Press the 

Enter

 key to confirm or the ESC key to cancel. 

 

 

 

Measurement and Calibration Hints 

 

Never pour the sample directly into the measurement chamber

 

and using a vial to perform measurement. 

 

Make sure the sample in vial is homogeneous. Do not shake or 

 

agitate the solution violently to prevent air bubbles. 

 

Keep the outside of vial clean and dry. If necessary, wipe the vial 

 

with a lint-free cloth. 

 

If the vial has scratches or scuffs, replace a new one (order code 

 

TB-GV). 

 

During the calibration and measurement processes, make sure 

 

the vial is positioned in the measurement chamber and the   

 

triangle mark on the vial must be aligned with the arrow on the 

 

meter. 

 

 

 

 

Always  close  the  light  shield  lid  to  prevent  the  measurement 

 

errors from ambient light. 

 

Wash the sample vial with distilled water after measurement. 

 

Any residues can cause inaccurate readings. 

 

 

 

Turbidity Calibration 

The TB200 turbidity meter allows turbidity calibration up to 7 points 
with minimum of 2 points, the default calibration points include the 
0.02, 10, 200, 500, 1000, 1500 and 2000 NTU. For better accuracy, we 
recommend to select a calibration point close to the sample value you 
are measuring. Note, the meter is calibrated with Formazin Standards 
at the factory and does not require user calibration before use. 

 

Calibrating the Meter 

1. 

Press the 

Cal

 key to enter the calibration mode. Press the 

 / 

 

 key to select the number of calibration points. 

 

 

 

 
2. 

Press the 

Enter

 key, the screen shows 0.02 NTU. If necessary, 

 

press the 

 / 

 key to select first calibration point, the meter 

 

will perform the calibration from the low to high turbidity. 

 

 

 

 
3. 

Take out the corresponding turbidity standard from accessory box 

 

(e.g., 0.02 NTU), wipe the vial with a lint-free cloth to remove 

 

fingerprints. Hold the vial cap and gently inverting vial several 

 

times, ensure that the turbidity standard is homogeneous and no

 

air bubbles. 

 

 

 
4. 

Insert

 

the vial into the measurement chamber.

 

Align the triangle 

 

mark on the vial cap with the arrow on the meter.

 

Close the light 

 

shield lid. 

 

 

 

Align arrows 

Mark 

Arrow 

Summary of Contents for TB200

Page 1: ...TB200 Laboratory Turbidity Meter USER MANUAL Bante Instruments Inc ...

Page 2: ...zin standard solutions 0 02 200 500 1000 NTU Two sample vials Lint free cloth 12V DC power adapter Meter Overview 1 Light shield lid 2 Measurement chamber 3 Display 4 Membrane keypad Connectors 1 Socket for power adapter 2 USB B interface to computer Keypad Key Function Switch the meter on or off Exit the calibration settings data logs and return to the measurement mode Start calibration Press and...

Page 3: ...calibration Enable Disable default Reset Reset the meter to factory default settings Note the meter must be recalibrated Enable Disable default Setting a Default Option 1 1 Press and hold the key to enter the setup menu 1 2 Press the key to select a menu item or an option 1 3 Press the Enter key to save To exit the setting without saving changes press the ESC key Setting the Password The password ...

Page 4: ...ow on the meter Always close the light shield lid to prevent the measurement errors from ambient light Wash the sample vial with distilled water after measurement Any residues can cause inaccurate readings Turbidity Calibration The TB200 turbidity meter allows turbidity calibration up to 7 points with minimum of 2 points the default calibration points include the 0 02 10 200 500 1000 1500 and 2000...

Page 5: ...n page 3 Measuring the Low Turbidity Sample For the low turbidity samples 200 NTU we recommend that using the same vial to perform the calibration and measurement 1 1 Rinsethe vial withapproximately10 mlof sample Cap andupend the vial several times Discard the used sample and repeat the rinsing procedure two more times 1 2 Fill the vial with the sample Cap the vial and wipe with a lint free cloth ...

Page 6: ...hem decrease to room temperature 3 Weigh the filter disk and watch glass and record the mg value as B 4 Filter the 100 ml of sample through pre weighed filter disk Put the filter disk and watch glass in a 104 C 219 F drying oven for 1 hour 5 Remove the filter disk and watch glass and put in a desiccator Immediately cover the desiccator Wait for them decrease to room temperature 6 Weigh the filter ...

Page 7: ...g the Data 1 Connect the USB cable to the meter and computer 2 Click the DAS icon the system automatically scans an available communication port and shows a message box Found a port on your computer 3 Click the OK the application starts 4 Click the Connect the screen shows Port is connected 5 Click the OK thenclick the Receive the storeddata will transfer to computer automatically If your computer...

Page 8: ...100 ml volumetric flask and dilute to the mark 200 NTU Mix 10 ml of 4000 NTU standard in a 100 ml volumetric flask and dilute to the mark Mix 50ml of above standard in a 100 ml volumetric flask and dilute to the mark 10 NTU Mix 2 5ml of 200 NTU standard in a 100ml volumetric flask and dilute to the mark Or Mix 2ml of 500 NTU standard in a 100ml volumetric flask and dilute to the mark Optional Acce...

Page 9: ...ter is one year from the date of shipment Above warranty does not cover the turbidity standards and glass vials Out of warranty products will be repaired on a charged basis The warranty on your meter shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the environment specifications of the products For mor...

Reviews: