Replacing the Electrode 

If the tester fails to calibrate or gives fluctuating readings, you should 
consider replacing the electrode. 
1. 

Twist the electrode collar counter clockwise, pull the electrode 

 

away from the tester. 

 

 

2. 

Align the slot on the new electrode, gently push the electrode 

 

into the tester. 

 

 

3. 

Twist the electrode collar clockwise until tight. 

 

 

 

 

 

Appendix 

 

Preparation of Conductivity Standard Solutions 

Place the analytical grade potassium chloride (KCl) in  a beaker and 
dry in an oven for about 3 hours at 105°C (221°F), then cool to room   
temperature. Add the reagent to a 1 liter volumetric flask according 
to the instructions in table below.   
 

Conductivity Standard 

Reagent 

Weight

84 µS/cm 

KCl 

42.35 mg

1413 µS/cm 

KCl 

745.5 mg

12.88 mS/cm 

KCl 

7.45 g

111.8 mS/cm 

KCl 

74.5 g

 
Fill the distilled water to the mark, mix the solution until the reagent 
is completely dissolved. 
 

Optional Accessories 

 
Order Code 

Description 

E-ECscan-C1-10K 

For SALscan10 tester, cell constant K=1, 
measuring range 10 to 20 mS/cm 

E-ECscan-C10-10K 

For SALscan20 tester, cell constant K=10, 
measuring range 0.1 to 200 mS/cm 

  
ECCS-84 

Standard solution 

84 μ

S/cm, 480 ml 

ECCS-1413 

Standard solution 

1413 μ

S/cm, 480 ml 

ECCS-1288 

Standard solution 12.88 mS/cm, 480 ml 

ECCS-50 

Standard solution 50 mS/cm, 480 ml 

ECCS-1118 

Standard solution 111.8 mS/cm, 480 ml 

 

Tester Specifications 

 
Model 

SALscan10 

SALscan20 

Conductivity 

 

Range 

0 to 20 mS/cm 

0 to 200 mS/cm 

Resolution 

0.01, 0.1, 1 

Accuracy 

±1% F.S. 

Calibration Points 

1 to 3 points 

Calibration Solutions 

84 

μ

S/cm 

1413 

μ

S/cm 

12.88 mS/cm 

1413 

μ

S/cm 

12.88 mS/cm 
111.8 mS/cm 

Salinity 

 

Range 

0 to 10 ppt 

0 to 80 ppt 

Resolution 

0.01 

Accuracy 

±1% F.S. 

Temperature 

 

Range 

0 to 60°C (32 to 140°F) 

Resolution 

0.1°C (0.1°F) 

Accuracy 

±1°C (±1.8°F) 

Calibration Point 

1 point 

Calibration Range 

Reading ±10°C/°F 

Other Specifications 

 

Temperature Compensation  0 to 60°C (32 to 140°F), automatic 

Temperature Coefficient 

2%/°C 

Reference Temperature 

25°C 

Cell Constant 

K=1 

    K=10 

Operating Temperature 

0 to 50°C (32 to 122°F) 

Storage Temperature 

0 to 60°C (32 to 140°F) 

Relative Humidity 

< 80% (non-condensing) 

IP Rating 

IP54 

Display 

LCD, 21 × 21 mm (0.82 × 0.82 in.) 

Power Requirements 

2 × 1.5V AAA alkaline batteries 

Auto-Off 

8 minutes after last key pressed 

Dimensions 

185 (L) × 40 (Dia.) mm (7.28 × 1.57 in.) 

Weight 

100

 

g (3.5 oz.) 

Summary of Contents for SALscan10

Page 1: ...SALscan10 20 Pocket Salinity Tester USER MANUAL Bante Instruments Inc ...

Page 2: ...tion Confirm the calibration settings or displayed option Toggle between salinity and conductivity measurement modes Display Icon Description When the battery voltage falls below the minimum power requirements the icon automatically disappears MEAS Indicates that the tester is in the measurement mode CAL Indicates that the tester is in the calibration mode SETUP Indicates that the tester is in the...

Page 3: ...hold the Cal key to enter the setup menu 2 If necessary press the Cal key again to select an option 3 Press the Enter key the tester saves the current option and moves to the next menu item 4 Repeat steps above until the tester returns to the measurement mode During the setting process if you do not need to calibrate the temperature press the Enter key to skip the or option To exit the setup menu ...

Page 4: ...When the setting is completed make sure that the displayed value matches the calibration standard 1 5 Press the Enter key the tester begins the calibration When the reading has stabilized the display will show Calibration is completed Multipoint Calibration Ensure that you have selected 2 or 3 points calibration in the setup menu 2 1 Repeat steps 1 1 through 1 5 above When the first calibration po...

Page 5: ...he sensor surface Wait for the measurement to stabilize and record the reading During the measurement DO NOTcompletely immerse thetester in water If the display shows indicating the measurement exceeds the range remove the tester from the sample immediately If the option is enabled in the setup menu the tester will automatically lock the measurement endpoint and show HOLD icon Press the Meas key t...

Page 6: ...SALscan20 tester cell constant K 10 measuring range 0 1 to 200 mS cm ECCS 84 Standard solution 84 μS cm 480 ml ECCS 1413 Standard solution 1413 μS cm 480 ml ECCS 1288 Standard solution 12 88 mS cm 480 ml ECCS 50 Standard solution 50 mS cm 480 ml ECCS 1118 Standard solution 111 8 mS cm 480 ml Tester Specifications Model SALscan10 SALscan20 Conductivity Range 0 to 20 mS cm 0 to 200 mS cm Resolution ...

Page 7: ...ester is one year from the date of shipment Above warranty does not cover the electrode and standard solutions Out of warranty products will be repaired on a charged basis The warranty on your tester shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the environment specifications of the products For mor...

Reviews: