background image

SALscan10/20   

I

    4 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
     
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

 

 
4. 

Place the electrode into a solution with a known accurate   

 

temperature and wait for the measurement to stabilize. 

 

 

 

 
5. 

Press the 

Cal

 key to modify the temperature value. 

 

 

 

 
6. 

Press the 

Enter

 key to save and press the 

Meas

 key to return 

 

to the measurement mode. 

 

 

 

 

 

 

Measurement 

 

Switching the Measurement Mode 

Press the 

Enter

 key, the tester will show

 

 

(conductivity),   

 

(salinity) and switch to the corresponding measurement mode   

automatically. 
 

 

 

 
Measuring the Sample 

Rinse the electrode with distilled water, place the electrode into the 
sample solution and stir gently, make sure that no air bubbles on the 
sensor surface. Wait for the measurement to stabilize and record the 
reading. 

 

 

 

During the measurement, DO NOT completely immerse the tester 

 

in water. 

 

     

 

 

 

If the display shows ---- indicating the measurement exceeds the 

 

range, remove the tester from the sample immediately. 

 

If the

 

 

option is enabled in the setup menu, the tester will 

 

automatically lock the measurement endpoint and show

 

HOLD 

 

icon. Press the 

Meas

 key to resume measuring. 

 

 

 

 

 

 

Electrode   
Maintenance and Replacement 

 

Rinse the electrode thoroughly with distilled water after use. 

 

Do not touch the platinum black coating on the sensor surface 

 

and always keep it clean. 

 

If there is a build-up of solids inside the sensor, remove carefully, 

 

then recalibrate the tester. 

 

If you do not use the tester for long periods, store the electrode 

 

with tap water. 

Platinum sensor 

Summary of Contents for SALscan10

Page 1: ...SALscan10 20 Pocket Salinity Tester USER MANUAL Bante Instruments Inc ...

Page 2: ...tion Confirm the calibration settings or displayed option Toggle between salinity and conductivity measurement modes Display Icon Description When the battery voltage falls below the minimum power requirements the icon automatically disappears MEAS Indicates that the tester is in the measurement mode CAL Indicates that the tester is in the calibration mode SETUP Indicates that the tester is in the...

Page 3: ...hold the Cal key to enter the setup menu 2 If necessary press the Cal key again to select an option 3 Press the Enter key the tester saves the current option and moves to the next menu item 4 Repeat steps above until the tester returns to the measurement mode During the setting process if you do not need to calibrate the temperature press the Enter key to skip the or option To exit the setup menu ...

Page 4: ...When the setting is completed make sure that the displayed value matches the calibration standard 1 5 Press the Enter key the tester begins the calibration When the reading has stabilized the display will show Calibration is completed Multipoint Calibration Ensure that you have selected 2 or 3 points calibration in the setup menu 2 1 Repeat steps 1 1 through 1 5 above When the first calibration po...

Page 5: ...he sensor surface Wait for the measurement to stabilize and record the reading During the measurement DO NOTcompletely immerse thetester in water If the display shows indicating the measurement exceeds the range remove the tester from the sample immediately If the option is enabled in the setup menu the tester will automatically lock the measurement endpoint and show HOLD icon Press the Meas key t...

Page 6: ...SALscan20 tester cell constant K 10 measuring range 0 1 to 200 mS cm ECCS 84 Standard solution 84 μS cm 480 ml ECCS 1413 Standard solution 1413 μS cm 480 ml ECCS 1288 Standard solution 12 88 mS cm 480 ml ECCS 50 Standard solution 50 mS cm 480 ml ECCS 1118 Standard solution 111 8 mS cm 480 ml Tester Specifications Model SALscan10 SALscan20 Conductivity Range 0 to 20 mS cm 0 to 200 mS cm Resolution ...

Page 7: ...ester is one year from the date of shipment Above warranty does not cover the electrode and standard solutions Out of warranty products will be repaired on a charged basis The warranty on your tester shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the environment specifications of the products For mor...

Reviews: