Temperature Compensation 

For better accuracy, we recommend to enable the manual temperature compensation for measurement and calibration. The tester will calculate 
the pH slope with inputted temperature and shows the temperature compensated readings. 
1.    Press and hold the 

Cal

 key to enter the setup menu. 

2. 

Press the 

Enter

 key until the display shows

 

/

 

 

or

 

/

 

 

3. 

Press the 

Cal

 key, the tester enters the temperature setting mode. 

4. 

Press the 

Cal

 key again to modify the temperature value. 

5. 

Press the 

Enter

 key to save and press the 

Meas

 key to return to the pH measurement mode. 

 

 

 

 

 

pH Calibration 

The

 

PHscan40 tester allows 1 to 3 points calibration, we recommend that you perform at least 2 points calibration for accurate measurement, 

the tester will automatically recognize and calibrate to following standard buffer values. 

 
USA standard buffers 

pH

 

4.01, 7.00, 10.01 

NIST standard buffers 

pH

 

4.01, 6.86, 9.18 

 

Single point calibration should only be carried out with pH

 

7.00 or 6.86, otherwise calibration will not be accepted. 

For better accuracy, we recommend calibrating the tester regularly. DO NOT reuse the buffer solutions after calibration, contaminants in solution 
will affect the calibration and eventually the accuracy of the measurement. 
 

Single Point Calibration 

Make sure that you have selected 1 point calibration in the setup menu. 
1.1  Press the 

Cal

 key, the tester shows "pH

 

7.00/CAL 1". 

1.2  Rinse the electrode with deionized water and place into the pH

 

7.00 buffer solution. Stir the electrode gently. 

1.3  Press the 

Enter

 key to begin the calibration. When the reading has stabilized, the display will show "End". Calibration is completed. 

 

 

 