PHscan40
Switching the Tester On and Off
•
Press and hold the
Meas
key for about 5 seconds to switch on the tester.
•
Press and hold the
Meas
key to switch off the tester.
•
To disable the auto-power off function, refer to the
Setup Menu
section.
Setup Menu
The PHscan40 tester contains 7 menu items in the setup menu, the following table describes the functions of each option.
Menu
Description
Options
Description
Default
Set the pH buffer group for calibration and auto-recognition
USA (pH
4.01, 7.00, 10.01)
USA
NIST (pH
4.01, 6.86, 9.18)
Set the number of calibration points
1, 2 or 3 points
2 points
Set the temperature unit
Degrees Celsius
° C
Degrees Fahrenheit
Set the temperature of sample
Range: 0 to 100° C
---
When the option is enabled, the tester will automatically
sense a stable reading and lock the measurement
Enable
Disable
Disable
When the option is enabled, the tester will automatically
switch off if no key is pressed within 8 minutes
Enable
Enable
Disable
Reset the tester to factory default settings
Enable
Disable
Disable
Setting the Default Option
1.
In the measurement mode, press and hold the
Cal
key for 5 seconds to enter the setup menu.
2.
If necessary, press the
Cal
key again to select an option.
3.
Press the
Enter
key, the tester saves the current option and moves to the next menu item.
4.
Repeat the steps above until the tester returns to the measurement mode.
•
During the setting process, press the
Meas
key, the tester will exit the setup menu and return to the measurement mode.
•
If you do not need to set the temperature of sample, press the
Enter
key to skip the
/
or
/
option.
•
The
option is used to restore the tester back to the factory default settings. If enabled, all of the calibration data and user-specific
settings will be deleted or reset, the tester must be recalibrated.
option