Bante Instruments Bante932 Instruction Manual Download Page 8

Bante932 Benchtop Water Hardness Meter 

 

Setup Menu 

The meter contains an integrated setup menu that is used to customize the displayed option to meet measurement requirements. The following table   
describes the functions of the menu items. 
 

MENU 

DESCRIPTION 

OPTIONS 

DESCRIPTION 

DEFAULT 

 

Calibration Points: 
Set the number of calibration points. 

 

2 points 

2 points 

 

3 points 

 

4 points 

 

5 points 

 

Stability Criteria: 
When the LO option is enabled, the Stable icon will   
quickly appear on the display. 
When the HI option is enabled, the icon will take   
longer to appear, but guarantees high accuracy of   
the measurement. 

 

Low 

Low 

 

High 

 

Auto-Hold: 
When the option is enabled, the meter will   
automatically sense a stable reading and lock the   
measurements. 

 

Enable 

Disable 

 

Disable 

 

Auto-Power Off: 
When the option is enabled, the meter will   
automatically turn off if no key is pressed within a   
specified time period. 

 

10 minutes 

Disable 

 

20 minutes 

 

30 minutes 

 

Disable 

 

Calibration Due: 
When the option is enabled, if the meter does not   
calibrated within a specified time period, the meter   
will automatically show the

 

 

icon. 

 

1 to 31 days 

Disable 

 

Disable 

 

Date and Time: 
Set the current date and time. 

--- 

Year-month-day, hour-minutes 

 

 

Clear Stored Data: 
Delete all stored readings in the memory. 

 

Enable 

Disable 

 

Disable 

 

Factory Reset: 
If enabled, all of the calibration data and selected   
parameters will back to factory default settings,   
the meter must be recalibrated. 

 

Enable 

Disable 

 

Disable 

 
 

 

Summary of Contents for Bante932

Page 1: ...Bante932 Benchtop Water Hardness Meter Instruction Manual BANTE INSTRUMENTS CO LTD...

Page 2: ...t work environment meets following conditions Relative humidity is less than 80 Ambient temperature is greater than 0 C and less than 60 C No potential electromagnetic interference The following list...

Page 3: ...meter is in the measurement mode Stable icon Indicates the measuring value has stabilized Calibration mode icon Indicates the meter is in the calibration mode Hold icon Indicates the measuring value...

Page 4: ...menu Press and hold the key for 3 seconds Stores current reading to memory Increase value or scroll up through the menu item Views the calibration report or data logs Decrease value or scroll down thr...

Page 5: ...tion Adjustment of electrode arm After installation if the electrode arm automatically rises or falls you need to adjust the screws until arm locate at any position 1 Remove the plastic cover from the...

Page 6: ...O NOT pull on the cable Always make sure that the connector is clean and dry Connecting the Temperature Probe 1 Place the temperature probe into the circular hole of the electrode arm 2 Insert the pho...

Page 7: ...Use Remove the protective cap from the bottom of the water hardness electrode Soak the electrode in the 0 01mol L standard solution for at least 30 minutes Switching the Meter On and Off Press the Me...

Page 8: ...Hold When the option is enabled the meter will automatically sense a stable reading and lock the measurements Enable Disable Disable Auto Power Off When the option is enabled the meter will automatica...

Page 9: ...ter returns to the measurement mode Setting is completed If you want to exit the setting press the Meas key Setting the date and time 2 1 Press and hold the key for 3 seconds to enter the setup menu a...

Page 10: ...meter is now switched to the automatic temperature compensation mode Manual Temperature Compensation If the meter does not detect a temperature probe the C icon will show on the display indicating tha...

Page 11: ...ted or calibration is not successfully the display will always show 0 000 During the calibration ensure that the selected calibration points cover the anticipated range of the samples Calibrating the...

Page 12: ...Stir the electrode gently Press the Enter key the Calibration icon begins flashing 1 7 Wait for the reading to stabilize the meter automatically show END and return to the measurement mode Calibratio...

Page 13: ...key until the meter shows ELE P 02 2 3 Press the Enter key the meter shows the last calibration date Format month day 2 4 Press the key the meter shows the first calibartion point and mV value e g 0...

Page 14: ...ardness Measurement Before measuring ensure that the temperature of samples are the same as the standard solutions the maximum error should be controlled within the 1 C For low level measurements or s...

Page 15: ...e display Viewing stored readings 1 Press the MR key in the measurement mode the meter shows LOC P 01 Data Log 2 Press the Enter key the meter shows the serial number of the stored data 3 Press the ke...

Page 16: ...shed Click the OK button to confirm Click the Receive button the stored data automatically transfer to computer Interval recording This function is used for recording the measuring value within the sp...

Page 17: ...ed value is out of range Check the electrode whether clogged dirty or broken Incorrect calibration solutions Using the fresh calibration solutions for calibration Electrode has expired Replace the ele...

Page 18: ...on 0 50 C Manual or Automatic Memory Stores up to 500 data sets Output USB communication interface Connector BNC Display LCD Operating Temperature 0 60 C Relative Humidity 80 Power Requirements DC5V u...

Page 19: ...to speed up these efforts Warranty The warranty period for meter is one year from the date of shipment Above warranty does not cover the sensor and calibration solutions Out of warranty products will...

Reviews: