Software Startup 

Please follow the user manual to 
install BWSpec. When the 
software is first opened set the 
workspace to Raman. 

 

 

 

Software Status Online 

Summary of Contents for i-Raman Plus

Page 1: ... 368 7824 Fax 1 302 368 7830 Web www bwtek com 290020251 C 2018 09 12 Page 1 of 16 Copyright 2018 B W Tek LLC i Raman Plus Quick Start Guide For full operating instructions please refer to the i Raman Plus portable system User Manual 290020245 ...

Page 2: ...4 Fax 1 302 368 7830 Web www bwtek com 290020251 C 2018 09 12 Page 2 of 16 Copyright 2018 B W Tek LLC Diagrams of the i Raman Plus Power Laser Emission Indicator Collection Port Laser Emission Port Flip Cover Trigger Port for auto collection function ...

Page 3: ...19 Shea Way Newark DE 19713 USA Tel 1 302 368 7824 Fax 1 302 368 7830 Web www bwtek com 290020251 C 2018 09 12 Page 3 of 16 Copyright 2018 B W Tek LLC ...

Page 4: ... Shea Way Newark DE 19713 USA Tel 1 302 368 7824 Fax 1 302 368 7830 Web www bwtek com 290020251 C 2018 09 12 Page 4 of 16 Copyright 2018 B W Tek LLC Distance Regulator Manual Shutter Trigger Button Shaft ...

Page 5: ... www bwtek com 290020251 C 2018 09 12 Page 5 of 16 Copyright 2018 B W Tek LLC Hardware Installation Connect probe to front panel Make sure FC connector to Laser Emission Port is in key in position DO NOT push the connectors into the ports too hard this will damage the optics ...

Page 6: ... 1 302 368 7824 Fax 1 302 368 7830 Web www bwtek com 290020251 C 2018 09 12 Page 6 of 16 Copyright 2018 B W Tek LLC Pull off distance regulator for liquid sample test with cuvette holder Use BAC150 stage for solid and powder sample tests ...

Page 7: ... 19713 USA Tel 1 302 368 7824 Fax 1 302 368 7830 Web www bwtek com 290020251 C 2018 09 12 Page 7 of 16 Copyright 2018 B W Tek LLC Testing powder through a plastic bag Direct contact measurement of pigments on a statue wall ...

Page 8: ...om 290020251 C 2018 09 12 Page 8 of 16 Copyright 2018 B W Tek LLC Connect Interlock Align two red dots in one line Gently push the plug into the receptacle until you hear a click sound which suggests the locking finger is now locked into the locking groove located inside the receptacle ...

Page 9: ...using NOT the cable or back nut which will cause the locking finger to disengage from the locking groove and the plug can be removed When the safety interlock plug is removed from the system all electrical power to the laser will be turned off The safety interlock MUST be inserted before the laser can be turned on Laser emission will stop if the interlock is removed while the laser is on ...

Page 10: ...4 Fax 1 302 368 7830 Web www bwtek com 290020251 C 2018 09 12 Page 10 of 16 Copyright 2018 B W Tek LLC Software Startup Please follow the user manual to install BWSpec When the software is first opened set the workspace to Raman Software Status Online ...

Page 11: ...7824 Fax 1 302 368 7830 Web www bwtek com 290020251 C 2018 09 12 Page 11 of 16 Copyright 2018 B W Tek LLC Software Interface Integration Time Integration Time Unit Multiplier Time Average Laser Power Percentage Scale Setting Display Reset Clear All ...

Page 12: ...4 Fax 1 302 368 7830 Web www bwtek com 290020251 C 2018 09 12 Page 12 of 16 Copyright 2018 B W Tek LLC Laser On Off Button Indicates laser is on Acquire One Spectrum Acquire Overlay Acquire Continuously Acquire Dark Scan Laser shuts off automatically ...

Page 13: ...8 7824 Fax 1 302 368 7830 Web www bwtek com 290020251 C 2018 09 12 Page 13 of 16 Copyright 2018 B W Tek LLC Spectrum x axis unit is set as Raman Shift Spectrum type is selected for display Raw Data or Dark Subtracted if dark scan is already taken ...

Page 14: ... 1 302 368 7830 Web www bwtek com 290020251 C 2018 09 12 Page 14 of 16 Copyright 2018 B W Tek LLC Before testing please confirm Power supply is on Interlock is plugged in Laser key is switched on Probe shutter is in open position Software status is online ...

Page 15: ...s around 50000 counts 2 Take dark scan and then take additional scan of the sample Be sure spectrum type is set to Dark Subtracted if dark scan is applied 3 Save data as txt or spc as preferred Warning Be sure to wear laser goggles at all times and avoid looking into laser beam Turn off fluorescent lamp or any ambient light which would bring noise signals into spectrum Lower the laser power if sam...

Page 16: ...cs Antioxidants Buffers Coatings Dilutants Emulsifiers Excipients Lubricants Preservatives Solvents Vitamins Monomers and Polymers Organic Materials Polyatomic Inorganics i e Calcium Phosphate Sodium Bicarbonate Titanium Dioxide Weak Raman Signal Given Materials that are Dark in Color Highly Fluorescing Molecules Fillers Binding Agents Thin Walled Plastics Water Glass No Raman Signal Given Black M...