Set-up
Fully remove transportation aids (pallets, transport safety devices).
Set up the ultrasonic/rinsing bath in a dry room.
Mount the separately-packaged ball valves for each tank as follows:
1. Screw the counter nut onto the threaded sleeves
1 at the tank outlet (3-way ball valve) and tank
over
fl
ow (ball valve), then seal with the white
PTFE tape provided, applying several turns of
tape to the right.
2. Tape the hose socket threads 2 with PTFE
sealing tape.
3. Screw the hose sockets 2 into the ball valve
outlets 3, use the appropriate spanner.
4. Screw the ball valve 3 onto the threaded
sleeves and tighten with the counter nut 4.
Caution:
Do not screw the ball valve up to the
limit stop or back it o
ff
by rotating it slightly to
the left, this will cause the PTFE tape to lose its
sealing e
ff
ect.
5. Close ball valve, turn the lever 5 upwards.
2.3 Start-up
Thoroughly rinse the ultrasonic or rinsing bath's oscillating tank with water before its
fi
rst use.
Note:
In order to protect the surface during transport and storage, all outer surfaces (and
also the inner walls of the oscillating tank) are covered with an oily preservative.
This should be removed with a suitable cleanser before
fi
rst use, see chapter 5.
Verify that the control buttons are in the "o
ff
" position, i.e. the switch indicator is at
the top, then connect the bath to the mains.
Conduct function test on the ultrasonic bath – brie
fl
y plug in the ultrasound
(maximum of 1 to 2 seconds), a hissing noise should be heard. Set to "0" once
again.
Hang accessories in the ultrasonic bath and place lid on top.
Connect peripheral devices if required – see additional documentation.
View from above:
4
1
3
2
5
open
open
closed
2
screwed-in