5.3. Treatment of medical instruments
- The instruments may not be placed on the tank bottom. The basket avoids damages
at the instruments and at the tank bottom. Do not staple instruments, overloading
reduces the cleaning efficiency. Pliers and scissors must be opened completely or
detached, if necessary. Instruments must be completely covered with cleaning liquid,
air has to escape from cavities and hoses.
- When disinfecting and cleaning contaminated medical instruments in ultrasonic baths,
the hygienic safety is important after the treatment. In case of improper and irregular
disinfection and cleaning a micro-biological contamination is possible caused by
settlement of microorganism, especially at the tank rim or in the drain section. Cross
infection might be a result. Therefore, the tank and housing surfaces must be regularly
disinfected and cleaned.
According to the hygienic plan the user must carry out regular disinfection and
cleaning with a surface disinfecting agent certified by DGHM*, or another efficiently
classified agent.
* Deutsche Gesellschaft für Hygiene und Mikrobiologie (German Society of Hygiene
and Microbiology)
6. Trouble
shooting?
•
Oscillations are too weak and non-uniform, the oscillation noise is too loud;
the cleaning effect is insufficient:
- Has the liquid been properly degassed?
⇒
sonicate for 15 minutes.
- Is the unit overloaded?
⇒
remove some parts.
- Non-uniform noise (wobbling) is not a defect
⇒
change the fluid level.
•
Heating defect:
No objection to operate the unit without the heating.
•
Minor erosion marks at the tank bottom?
⇒
common wear and tear effect.
The unit is alright.
•
The unit is equipped with an overvoltage arrestor.
It protects the unit from voltage surges
⇒
mains fuse can be tripped.
•
Repairs should only be carried out by authorised service.
•
Replace defective parts only by original
SONOREX
parts.
- 19 -