5. Recommendations 

 

5.1. Cleaning agents 

 

 

Best cleaning results are achieved with TICKOPUR, TICKOMED and STAMMOPUR. 

 

Replace used liquids and disinfectants, do not "freshen them up" with an added dose.  

 

Layers of dirt on the tank bottom or in the insets (abrasives and polishing agents, for 
example, as well as other deposits) reduce the cleaning efficiency. Empty the 
oscillating tank and clean it regularly. Replace the cleaning liquid.  

 

The cleaning and contact liquids in the ultrasonic oscillating tank may not exceed a 
maximum operating temperature of 100 °C. 

 

Distilled or deionized water without additives is unsuitable for ultrasonic cleaning 
directly in the oscillating tank. They lack the necessary cavitation properties to form a 
uniform ultrasonic field. Irregular cavitation can damage the oscillating tank or the 
objects being cleaned. (You may use distilled or deionized water in the inset beakers 
and non-perforated insert tubs.) 

 

Use only TICKOPUR, STAMMOPUR (see above) in the oscillating tank. These agents 
are especially developed and recommended for ultrasonic cleaning.  

 

Be careful when working with aggressive cleaning agents in insets or non-perforated 
insert tanks; avoid splashing into the contact liquid or onto stainless steel surfaces; if 
necessary, replace the contact liquid immediately, clean the surfaces and wipe them 
dry.  

 
 

The manufacturer does not take any responsibility for equipment damage caused 
by unsuitable cleaning chemicals. 

 

 

SONOREX

 

units do not require maintenance.

 

 
 
 

5.2.  Cleaning and care  

 

-  Be sure to pull the plug before cleaning the unit. 
- Do 

not shower the units or immerse them in water; instead, wipe them with a damp 

cloth, in- and outside. 

-  Rinse out the stainless steel tank frequently and wipe it dry. 
-  Do not use steel wool, scrapers or graters. 
-  If rings should appear in the stainless steel tank after a longer period of operation, 

remove them with a customary stainless steel cleaner without scouring additives.  

-  Metal parts remaining on the stainless steel surface, as well as particles of rust, can 

penetrate the passive protective layer of the stainless steel, thereby "activating" it and 
causing it to rust. This rust provokes pittings on the stainless steel surface. Therefore, 
remove all remaining metal parts such as screws, metal chips and the like; remove 
minor rust spots immediately with a soft cloth and customary stainless steel cleaners 
without scouring additives. 

- 18 - 

Summary of Contents for Sonorex Longlife

Page 1: ...If service is required please contact your dealer or the address mentioned below Mode d emploi 3 6 et 3 0 Appareils de nettoyage par ultrasons pour solutions de nettoyage aqueuses Appareils de nettoya...

Page 2: ...Ein Aus 6 3 4 Ger te mit Heizung 6 3 5 Schwingwanne entleeren 7 4 Montage und Demontage der Ablaufgarnitur TK 100 H 7 5 Weitere Empfehlungen 5 1 Reinigungsmittel 8 5 2 Reinigung Pflege des Ger tes 8...

Page 3: ...ingt beachten Nach dem Entfernen der Verpackung Geh use und Bedienelemente des Ger tes auf eventuelle Transportsch den berpr fen Falls Transportsch den festgestellt werden Ger t nicht an das Netz ansc...

Page 4: ...ase z B Sauerstoff Frisch eingef llte oder l ngere Zeit in der Schwingwanne verbliebene Reinigungs oder Kontaktfl ssigkeit vor Gebrauch ca 5 bis 15 min beschallen Bei mehr als 10 l Volumen etwa 30 min...

Page 5: ...schutz Richtlinien EX RL BGR 104 in der jeweils g ltigen Fassung zu beachten Einh ngewanne aus Kunststoff Reinigungs fl ssigkeit Kontakt fl ssigkeit 2 3 Einsatzgef e in Lochdeckel einh ngen nicht auf...

Page 6: ...dauer ca 1 bis 15 min 3 4 Ger te mit Heizung Ger te mit dem Buchstaben H in der Bezeichnung besitzen eine Heizung Die Heizung arbeitet unabh ngig vom Ultraschall 0 RK 31 H und TK 100 H Wippenschalter...

Page 7: ...folgter Quetschung nicht gleich ffnet Ger te ohne Ablauf Wanneninhalt ber Eck ausgie en dabei Spritzer von unten an das Ger t vermeiden 4 Montage und Demontage der Ablaufgarnitur TK 100 H Ablaufgarnit...

Page 8: ...keit oder auf die Edelstahlfl chen ver meiden ggf Kontaktfl ssigkeit sofort erneuern Fl chen s ubern und trocken reiben Bei Ger te und Instrumentensch den verursacht durch Anwendung ungeeigneter Reini...

Page 9: ...en kann Deshalb ist die Wannen und Ger teoberfl che regelm ig zu desinfizieren und zu reinigen Die Desinfektion und Reinigung ist durch den Betreiber regelm ig und entsprechend dem Hygieneplan mit ein...

Page 10: ...ml 6 4 l 200 ml 6 2 l 330 ml 5 9 l 660 ml RK 512 H CH 8 7 l 8 6 l 90 ml 8 5 l 180 ml 8 4 l 270 ml 8 2 l 440 ml 7 8 l 870 ml RK 513 12 0 l 11 8 l 120 ml 11 7 l 240 ml 11 6 l 360 ml 11 4 l 600 ml 10 8...

Page 11: ...ls Medizinprodukt Bezeichnung Ultraschall Reinigungsger t UMDNS Nomenklatur ECRI DIMDI 14 263 Zweckbestimmung Ultraschall Desinfektion und Reinigung medizinischer und zahnmedizinischer Instrumente Kla...

Page 12: ...FF 16 3 4 Units with heating 16 3 5 Emptying the ultrasonic tank 17 4 Fixing and removing of the outlet at TK 100 H 17 5 Additional recommendations 5 1 Cleaning agents 18 5 2 Cleaning and care 18 5 3...

Page 13: ...n cleaning temperature sensitive parts 2 Safety instructions to be carefully observed After removing the packing check the unit s casing and control elements for any signs of damage In such case do no...

Page 14: ...st been poured into the oscillating tank or have been left in it for a longer period of time for approx 5 to 15 minutes before starting the cleaning process Degas oscillating tanks with a volume of mo...

Page 15: ...spend inset beakers in the positioning cover do not place them on the tank bottom Insert plastic tub Cleaning liquid Contact liquid 2 3 2cm 2 3 Inset beakers or insert plastic tub must be immersed in...

Page 16: ...ve a built in heating The heating system runs independently from the ultrasonic energy RK 31 H and TK 100 H Tip switch Temp 65 C 0 Tip switch operation fixed temperature of 65 C Heating switch on cont...

Page 17: ...clamp position does not open immediately caused by quenching Units without drain Pour out the liquid over one of the tank corners Avoid splashing from below 4 Fixing and removing of the outlet at TK...

Page 18: ...th aggressive cleaning agents in insets or non perforated insert tanks avoid splashing into the contact liquid or onto stainless steel surfaces if necessary replace the contact liquid immediately clea...

Page 19: ...must be regularly disinfected and cleaned According to the hygienic plan the user must carry out regular disinfection and cleaning with a surface disinfecting agent certified by DGHM or another effic...

Page 20: ...0 ml RK 512 H CH 8 7 l 8 6 l 90 ml 8 5 l 180 ml 8 4 l 270 ml 8 2 l 440 ml 7 8 l 870 ml RK 513 12 0 l 11 8 l 120 ml 11 7 l 240 ml 11 6 l 360 ml 11 4 l 600 ml 10 8 l 1 2 l RK 514 H 9 0 l 8 9 l 90 ml 8 8...

Page 21: ...C permissible relative humidity up to 31 C 80 permissible relative humidity up to 40 C 50 Dew not permitted Information for the use as a medical product Denotation Ultrasonic cleaning device UMDNS no...

Page 22: ...areils avec chauffage 26 3 5 Vidange de la cuve de vibration 27 4 Montage et d montage de la garniture d coulement TK 100 H 27 5 Autres recommandations 5 1 Nettoyants 28 5 2 Nettoyage et entretien de...

Page 23: ...ivement Apr s avoir enlev l emballage il convient de v rifier si le bo tier et les l ments de commande de l appareil n ont pas subi de dommages ventuels lors du transport En cas d un dommage ne pas br...

Page 24: ...ement remplies ou tant rest es depuis longtemps dans la cuve de vibration doivent tre trait es aux ultrasons pendant 5 15 minutes Les cuves de vibration de plus de 10 l de volume doivent tre d gaz es...

Page 25: ...olution de nettoyage Liquide de contact 2 3 Suspendre les b chers d insertion dans le couvercle perfor ne pas les d poser sur le fond de la cuve Les b chers ou la cuve suspendre doivent tre immerg s a...

Page 26: ...un chauffage Le chauffage fonctionne ind pendamment de l ultrason RK 31 H et TK 100 H interrupteur bascule Temp 65 C 0 Interrupteur bascule temp rature 65 C Interrupteur bascule de chauffage en marche...

Page 27: ...le pince il est possible que le tuyau n ouvre pas imm diatement Des appareils sans vidange Vider le contenu de la cuve sur le coin Eviter des claboussures de dessous sur l appareil 4 Montage et d mont...

Page 28: ...ou sur l acier inox le cas ch ant la solution de contact doit tre imm diatement remplac e les surfaces doivent tre nettoy es et frott es jusqu ce qu elles soient s ches Le fabricant se d gage de tout...

Page 29: ...es de la cuve et du bo tier La d sinfection et le nettoyage doivent tre effectu s par l applicateur r guli rement et selon le plan hygi nique avec un produit certifi par DGHM ou un autre d sinfectant...

Page 30: ...270 ml 8 2 l 440 ml 7 8 l 870 ml RK 513 12 0 l 11 8 l 120 ml 11 7 l 240 ml 11 6 l 360 ml 11 4 l 600 ml 10 8 l 1 2 l RK 514 H 9 0 l 8 9 l 90 ml 8 8 l 180 ml 8 7 l 270 ml 8 5 l 450 ml 8 1 l 900 ml RK 5...

Page 31: ...ible jusqu 31 C 80 humidit relative admissible jusqu 40 C 50 Condensation pas admissible Indications pour l application comme produit m dical D signation Cuve ultrasons Nomenclature UMDNS ECRI 14 263...

Page 32: ...K 170 182 1000 200 200 26 0 b 1050 250 385 27 0 2 600 1 4 F4A RK 170 H 076 1000 200 200 26 0 b 1050 250 385 28 5 2 600 1600 8 3 F4A M15A RK 255 H 316 300 150 150 3 8 b 325 175 305 6 1 2 320 280 2 0 F3...

Reviews: