background image

Contact 

fl

 uid

Laboratory 

fl

 asks with 

sample 

fl

 uid

 roller carriage

1

2

3

18 / 25

51336-005 en/2022-03

1

 electronic GmbH & Co. KG  •  Heinrichstraße 3-4  •  12207 Berlin  •  Deutschland  •  [email protected]

4 Use 

 

The  shaking device is used together with the SONOREX DIGITEC DT 1028 F 

fl

 at 

 ultrasonic bath. Depending on the application, the shaking usually occurs indirectly in 
laboratory 

fl

 asks or other vessels, which are positioned in the 

 roller carriage.
Direct sonication of objects in the bath 

fl

 uid is not permitted in 

the DT...F ultrasonic baths and the  shaking device.

Notes on the use with the ultrasonic coupling

If inset beakers are used in the  ultrasonic bath, a contact 

fl

 uid 

must always be used as otherwise the ultrasound cannot pass 
through the air (between the ultrasonic oscillating tank and the inset beaker). This 
contact 

fl

 uid must be able to transfer the ultrasound to the 

fl

 uid in the inset beaker 

without loss if possible.
For more detailed information and instructions on correct ultrasound applications, 
please refer to the user instructions for the  ultrasonic bath.

4.1  

Preparing the  roller carriage for the application  

To avoid the 

fl

 asks' moving around or tipping over during operation of the  shaking 

device, BANDELIN recommends the use of special "EK"  holding clamps, see Section 7. 

The device must always be switched o

  before the  roller carriage is equipped with the 

 holding clamps. The  ultrasonic bath should also be switched o

  during the procedure. 

 Release the  coupling on the  roller carriage: Grasp the lever 1 behind the  bearing cup 

2 and pull it upwards. Leave or place the  clip bearing 3 back on the  ball stud. The 
 roller carriage can now be taken o

  the ultrasonic bath. 

 Equip the  roller carriage with the  holding clamps or replace the clamps if already 

mounted. 

 Position the  roller carriage back on the rim of the bath and check it for free 

movement.

 Press/click the  bearing cup 2 onto the lever 1 on the  ball stud (with  clip bearing 3). 

Summary of Contents for Sonorex Digitec DT 1028 F

Page 1: ...User Instructions SONOSHAKE Shaking device SA 1028 Accessories for SONOREX DIGITEC DT 1028 F ultrasonic baths 51336 005en 2022 03...

Page 2: ...en the translation and the original of this document the original version has priority BANDELIN accepts no responsibility or liability for damages caused by improper handling or usage contrary to the...

Page 3: ...t of unauthorised alterations modifications both the warranty claim and the CE conformity will expire If any service is required please contact the specialist dealer in charge or the manufacturer Symb...

Page 4: ...ectromagnetic ambient conditions EMC 9 1 5 Warnings and safety precautions 10 2 Preparation 11 2 1 Scope of delivery 11 2 2 Set up assembly 12 2 2 1 Fastening material set 12 2 2 2 Installing the moun...

Page 5: ...e for the application 18 4 2 Shaking 19 5 Cleaning disinfection and care of the device 21 5 1 Warehousing storing 21 6 Maintenance and repair 22 6 1 Repairs and service 22 6 2 Decontamination certific...

Page 6: ...continuous operation 3 Knob power selection switch for 0 25 50 75 or 100 4 Drive unit with compact easy to clean stainless steel housing 5 1 1 Mode of operation The shaking device comprises a roller...

Page 7: ...f fluids with a high gas content For the intended use the samples must be placed in suitable vessels e g laboratory flasks and secured in the roller carriage indirect operation Existing SONOREX DIGITE...

Page 8: ...z mains cable length 2 m Max current consumption 0 8 A 230 V 1 3 A 115 V Protection class I Emission sound pressure level A weighted 70 dB A Drive unit Exterior dimensions 225 320 235 mm Weight 4 kg i...

Page 9: ...tic ambient conditions EMC The device was tested to DIN EN 61326 1 for electromagnetic compatibility EMC and complies with the requirements for class B devices according to EN 55011 It is suitable for...

Page 10: ...nd disinfection only use cleaning agents and surface disinfectants which are suitable approved tested and compatible with the materials The user instructions with their safety instructions must remain...

Page 11: ...arriage separately Do not allow the unit to run without supervision 2 Preparation Carefully unpack the shaking device and accessories and inspect them for completeness or possible transportation damag...

Page 12: ...bH Co KG Heinrichstra e 3 4 12207 Berlin Deutschland info bandelin com 2 2 Set up assembly Danger Guard against moisture and wetness risk of electric shock Installation of the shaking device SA 1028 i...

Page 13: ...right handle of the ultrasonic bath Place the terminal strip on the handle Secure the mount and the terminal strip on the handle with the washers and screws Important The mount must be secured central...

Page 14: ...ly on the rim of the bath If the roller carriage wobbles or grinds on the rim of the bath it should be adjusted using the screw rear left outside After adjustment check whether the clamp bearing is on...

Page 15: ...with the mains 2 4 Start up Release the coupling on the roller carriage Grasp the lever 1 behind the bearing cup 2 and pull it upwards Leave or place the clip bearing 3 back on the ball stud The rolle...

Page 16: ...utes Once the time has elapsed the time switch will automatically turn o By turning the knob back the operating time will be shortened i e the time switch will be turned o Continuous operation Turn kn...

Page 17: ...3 2 Power setting The shaking frequency will be adjust via the knob for power setting shakes min Possible power settings shaking frequency shakes min 0 25 50 75 100 0 55 110 165 220 The shaking freque...

Page 18: ...fluid in the inset beaker without loss if possible For more detailed information and instructions on correct ultrasound applications please refer to the user instructions for the ultrasonic bath 4 1...

Page 19: ...laced The laboratory flasks can be secured in the roller carriage with EK holding clamps see chapter 7 1 Equip the roller carriage with laboratory flasks The laboratory flasks must be immersed at leas...

Page 20: ...flasks from the roller carriage CAUTION Depending on the length of the sonication the flasks samples may be hot Before the next sonication verify the service life of the contact liquid Heed the speci...

Page 21: ...aneous rust produces localised corrosion of the roller carriage For this reason remove rust stains immediately using a soft cloth and a commercial stainless steel cleaning product without abrasive add...

Page 22: ...88 02 00 13 In the case of returns the general terms and conditions for delivery and payment of BANDELIN electronic GmbH Co KG shall apply In addition the device is to be cleaned and decontaminated i...

Page 23: ...s or from our website No obligation telephone consultation Internet 49 0 30 768 80 0 www bandelin com 7 1 Required accessories EK holding clamps are required for securing the laboratory flasks They pr...

Page 24: ...ste Electrical and Electronic Equipment Directive 2012 19 EU Any supplementary deviating regulations must be observed The device must be decontaminated before disposal It can then be disposed of as el...

Page 25: ...Note The user instructions in this and other languages as well as further information are available on the supplied CD...

Reviews: