PRECAUTIONS FOR ACCURATE MEASUREMENTS

• The sample must not content any particles. 

• The test tube must be dry on the outside without finger prints or oil or

impurities. If necessary, clean it and wipe it dry - do not use abrasive
cleaners. 

• Shaking with the reagent may cause gas bubbles (therefore a highest

measured value). Before to place the test tube inside the tube holder,
move gently the test tube until all gas bubbles disappear. 

• Do not wait between end of reagent mixing and measurement in order

to warranty the highest possible accuracy. 

• Empty, clean, rinse and dry the test tube, as soon as possible, in order

not to colour it and decrease the quality of next test. 

DISPLAY OF ERROR MESSAGES

Light High:

Too much transmitted light.

Check the zero operation

Light Low:

Too weak transmitted light. 

Check the zero operation

Inverted test tubes: 

Test tubes, on zero and test,

have been exchanged.
Repeat the test operation. 

Under range:

Blinking display 

“0.00”

shows a

sample with less absorption than the zero test. 
Repeat the zero operation. 

Over range:

Blinking display 

“2.50”

indicates an 

over range of sample. 
Proceed to dilute carefully the sample and proceed
to a new test. 

Battery low:

Change the battery as soon as possible. 

Dead battery:

Change immediately the battery

With such a displayed message, all running test

parameters are lost. 

BATTERY LIFE TIME

• The meter automatically turns off after 2 minutes of non-use or 10 seconds

after measurement display.

One fresh battery lasts for about 5000 measurements, depending on
the light level. When the battery capacity is lower than 10% 

"bAt" 

appears on the LCD at start-up. 

If the battery is empty and accurate measurement can't be done any
more, the instrument shows 

“bAd”

then 

“bAt”

each for 1 second and

turns off. To restart the instrument, replace the battery by a fresh one. 

BATTERY REPLACEMENT

• Turn the instrument off, if not yet, by holding the button until the meter

shuts off. 

• Turn the instrument upside down and remove the battery cover with a

screw driver.

• Remove the battery from its location and replace it by a fresh one. 

• Insert the battery cover and replace the screw with a screwdriver.

RECOMMENDATIONS

Before  to  use  this  instrument,  verify  the  correspondence  between  the
range to be measured and the features of the instrument. 

The use of these photometers could interfere with other electronic devices. 

Any modification caused by the user, or unconsidered ambient conditions,
may affect the instrument. 

To avoid injury, do not put the instrument in a microwave oven. 

For  safety  reasons,  do  not  introduce  the  instrument  in  a  dangerous,
explosive area. 

Waste Electrical and Electronic Equipment (WEEE):

To  avoid  harm  to  the  environment  and  natural  resources,  the  use  of
appropriate recycling systems is recommended. 

For any requirement, please contact 

BAMO Mesures SAS

Reviews: