ADJUST ALARM 3
OUTPUT mA TURB
ALARM 3
ON/OFF
ALARM 3
MEASURE/TEMP
MEASURE
= Threshold against pH/mV measured value
TEMP.
= Threshold against temperature measured value
ALARM 3 LOW/HIGH
HIGH
= Contact closes when value goes over the limit
LOW
= Contact closes when the value goes under the limit
ON
0000
NTU / °C
To close the contact S3 at this value
OFF
0000
NTU / °C
To open the contact S3 at this value
DELAY UP
ONI/OFF
Delay (or no delay) before to close the contact S3
TIME
0000
SEC
Delay time to close the contact S3
DELAYDOWN
ON/OFF
Delay (or no delay) before to open the contact S3
TIME
0000
SEC
Delay time to open the contact S3
SAVING ?
OUTPUT mA TURBIDITE
OUTPUT mA TEMP
page 12
HIGHER
0000
NTU
Value corresponding to 20,00 mA
LOWER
0000
NTU
Value corresponding to 00,00 mA or 04,00 mA
OUTPUT
4/20 mA / 0/20 mA
Output type
SAVING ?