background image

436 M1 01 C

Page

10

ADJUST ALARM 1

ADJUST ALARM 2

ALARM 1 

ON/OFF

ALARM 1 

MEASURE/TEMP

MEASURE

= Threshold against pH/mV measured value 

TEMP.

= Threshold against temperature measured value 

ALARM 1 LOW/HIGH

HIGH

= Contact closes when value goes over the limit

LOW

= Contact closes when the value goes under the limit

ON 

0000

NTU / °C

To close the contact S1 at this value

OFF 

0000

NTU / °C

To open the contact S1 at this value

DELAY UP 

ONI/OFF

Delay (or no delay) before to close the contact S1

TIME 

0000

SEC

Delay time to close the contact S1

DELAYDOWN 

ON/OFF

Delay (or no delay) before to open the contact S1

TIME 

0000

SEC

Delay time to open the contact S1

SAVING ?

ADJUST ALARM 2

ADJUST ALARM 3         page 11

ALARM 2 

ON/OFF

ALARM 2 

MEASURE/TEMP

MEASURE

= Threshold against pH/mV measured value 

TEMP.

= Threshold against temperature measured value 

ALARM 2 LOW/HIGH

HIGH

= Contact closes when value goes over the limit

LOW

= Contact closes when the value goes under the limit

ON 

0000

NTU / °C

To close the contact S2 at this value

OFF 

0000

NTU / °C

To open the contact S2 at this value

DELAY UP 

ONI/OFF

Delay (or no delay) before to close the contact S2

TIME 

0000

SEC

Delay time to close the contact S2

DELAYDOWN 

ON/OFF

Delay (or no delay) before to open the contact S2

TIME 

0000

SEC

Delay time to open the contact S2

SAVING ?

L

L

L

L

L

L

L

L

L

L

L

L

L

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

VALID

MENU

+

L

L

L

L

L

L

L

L

L

L

L

L

L

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

VALID

MENU

+

Summary of Contents for BAMOPHOX 436 E

Page 1: ...bamo fr Fax 33 01 34 10 16 05 E mail info bamo fr Turbidity monitor and controller BAMOPHOX 436 MES 436 01 1 INSTRUCTION MANUAL INSTRUCTION MANUAL B B B BA A A AM M M MO O O OP P P PH H H HO O O OX X...

Page 2: ...ontents 1 TECHNICAL SPECIFICATIONS Page 3 2 CAUTION 3 3 DIMENSIONS 3 4 WIRING 4 5 FRONT PANEL 7 SCROLLING MENU 8 ABOUT BAMOPHOX 8 CONSULTING MODIFYING 8 MEASUREMENT PARAMETERS 9 ADJUSTING THRESHOLD 1...

Page 3: ...supply 230 V AC 50 60 Hz other on request Consumption 10 VA Models Panel mounting front IP65 72 x 144 mm connections on screw terminal IP40 Idem DIN Rail mounting only for blind monitor IP40 Wall mou...

Page 4: ...Page 4 436 M1 01 C 4 WIRING WALL MOUNTING OPTION LOGGER RS 422 OPTION LOGGER access by the upper cover RS 422 PANEL MOUNTING Measurement input...

Page 5: ...ld NO contact Measurement Temperature S3 threshold NO contact Measurement Temperature NOT USED Ground N Main power supply N Neutral L Line L NC mA Turbidity transmitter mA 1 2 3 4 5 6 7 8 9 10 11 12 1...

Page 6: ...blind BAMOPHOX B A S4 contact NO alarm on temperature or over range S2 threshold NO contact Measurement Temperature S1 threshold NO contact Measurement Temperature S3 threshold NO contact Measurement...

Page 7: ...key all parameters are saved previous data programmation will be overwrited If you are not sure of any modification do not press the VALID key To change parameters of data capture Num ric input increa...

Page 8: ...MENU MENU MENU MENU MENU MENU MENU MENU MENU ABOUT Bamophox ABOUT Bamophox VERSION 2 04 SERIAL N 20451 05 CONSULTATION MODIFICATION CONSULTATION MODIFICATION CODE 0000 CODE 5105 Last 4 digits of seria...

Page 9: ...NTU Decimal point selection SCALE MINI 00 00 NTU Minimal range value MAXI 00 00 NTU Maximal range value TRANSMITTER MINI 04 00 mA Minimal calculated value to display MAXI 20 00 mA Maximal calculated v...

Page 10: ...1 SAVING ADJUST ALARM 2 ADJUST ALARM 3 page 11 ALARM 2 ON OFF ALARM 2 MEASURE TEMP MEASURE Threshold against pH mV measured value TEMP Threshold against temperature measured value ALARM 2 LOW HIGH HIG...

Page 11: ...DELAY UP ONI OFF Delay or no delay before to close the contact S3 TIME 0000 SEC Delay time to close the contact S3 DELAYDOWN ON OFF Delay or no delay before to open the contact S3 TIME 0000 SEC Delay...

Page 12: ...rresponding to 00 00 mA or 04 00 mA OUTPUT 4 20 mA 0 20 mA Output type SAVING L L L L L ENTER ENTER ENTER ENTER VALID MENU FORCED RELAY LANGUAGE ALARM 1 ON OFF ALARM 2 ON OFF Diagnostic mode to test t...

Reviews: