www.balluff.com
BIS V-6108 PROFINET
Processor Unit
59
english
Command Identifier 50
hex
: Kill
The
Kill
command deactivates a data carrier previously selected with the Select command.
Note
Executing the Kill command permanently deactivates the selected data
carrier.
The deactivation cannot be undone.
Note
In order to execute the Kill command, a Kill password must first be set and written
to the data carrier.
Information about password protection and about locking and unlocking ("Lock")
of UHF RFID data carriers can be found in the UHF RFID standards EPCglobal™
Radio Frequency Identity Protocols Class-1 Generation-2 UHF RFID and
ISO IEC 18000-63.
The EPCglobal™ standard is available online at
www.gs1.org/standards
.
Subaddress
Meaning
Description of Function
00
hex
1st Bit Header
01
hex
Command Identifier 50
hex
: Kill
02
hex
Password 1
1st byte password
03
hex
Password 2
2nd byte password
04
hex
Password 3
3rd byte password
Password 4
(High Byte)
4th byte password
…
None
No meaning
Last byte
2nd Bit Header
If the 1st and 2nd bit headers match, the data is valid.
or
Input Buffer: Status Message
Subaddress
Meaning
Description of Function
00
hex
1st Bit Header
01
hex
Status code
Provides information on the status of a query.
…
None
No meaning
Last byte
2nd Bit Header
If the 1st and 2nd bit headers match, the data is valid.
Specific
commands for
BIS VU read/write
heads
7