Place the required items on the interlock tray. The weight limit for the interlock floor tray is
4kg.
•
Close the interlock outer door.
•
Start the interlock purge cycle by pressing the interlock cycle start button (item 2 in Figure
13)
•
Access the workstation chamber via the Ezee Sleeves. See section 6.2 for details.
•
When the interlock purge cycle has completed, indicated by the interlock cycle status
anaerobic (item 12 in Figure 9 or item 20 in Figure 10) being illuminated, press the inner
door button (item 2 in Figure 12).
•
Slide the interlock inner door backwards to open the interlock door.
•
Slide the interlock floor tray into the workstation chamber by pulling it to the left.
•
Move the material from the interlock chamber into the workstation chamber.
•
Slide the interlock floor tray into the interlock chamber by pushing it right.
•
Close the interlock inner door by sliding the interlock inner door forward.
Removing material from the workstation chamber via the interlock
6.1.5
To remove material from the workstation chamber via the interlock;
•
Open the interlock inner door by pressing the inner door button (item 2 in Figure 12) and
sliding the interlock inner door backwards.
•
Slide the interlock floor tray into the workstation chamber by pulling it to the left.
•
Move the material from the workstation chamber into the interlock chamber.
•
Slide the interlock floor tray into the interlock chamber by pushing it right.
•
Close the interlock inner door by sliding the interlock inner door forwards.
•
Exit the workstation chamber via the Ezee Sleeves.
•
Open the interlock outer door.
•
Remove the material from the interlock chamber.
•
Close the interlock outer door as described in section 6.1.2
Hand access to the main chamber
6.2
Direct hand access to the workstation chamber is provided via the glove ports and Ezee Sleeve
system.
The glove ports and Ezee Sleeves can be used in 3 ways;
•
Left hand glove port only, for loading items into the workstation chamber via the Single Plate
Entry System (SPES).
•
Both glove ports, for working inside the workstation chamber, loading items into the
workstation chamber via the SPES and unloading the interlock.
•
Right hand glove port only, for loading / unloading the interlock.
Ezeeyin Sleeve System Overview
6.2.1
The Ezeeyin system consists of Ezee Sleeve and a cuff.
•
The sleeve attaches to the glove port via 2 O-rings.