To replace the detox sachets via the interlock;
•
Remove the new detox sachets from their packaging.
•
Open the interlock outer door (see section 6.1.2, for more details).
•
Place the new detox sachets in the interlock chamber.
•
Close the interlock outer door (see section 6.1.2, for more details).
•
Start the interlock purge cycle by pressing the interlock cycle start button (item 2 in Figure
13).
•
Enter the workstation chamber via the Ezee Sleeves (see section 6.2 for more details).
•
Move any samples/ equipment onto the rear shelf. Do not exceed the weight limit for the
rear shelf.
•
Lift the front of the floor tray to expose the old detox sachets and sachet holders.
•
Whilst holding the floor tray up, slide the old detox sachets forwards and remove from
sachet holders.
•
When the interlock purge cycle has completed, open the interlock inner door (see section
6.1.3 for more details).
•
Place the old detox sachets in the interlock chamber and bring the new detox sachets into
the workstation chamber.
•
Close the interlock inner door (see section 6.1.3 for more details).
•
Whilst lifting the front of the floor tray, slide the new detox sachets into the detox sachet
holders.
•
Lower the front of the floor tray back to its original position.
•
Move any samples/ equipment back to the floor tray.
•
Exit the workstation chamber via the Ezee Sleeves (see section 6.2 for more details).
•
Open the interlock outer door (see section 6.1.2 for more details).
•
Remove the old detox sachets.
•
Close the interlock outer door (see section 6.1.2 for more details).
To remove the detox sachets with the front screen removed;
•
Remove all cells/ samples to an alternative storage facility.
•
Switch the workstation off at the mains and remove the plug from the mains.
•
Remove the Ezee Sleeves (see section 7.4.2 for more details).
•
Remove any other equipment installed in the workstation chamber.
•
Lift the front of the floor tray to expose the detox sachets.
•
Whilst holding the floor tray up, slide the detox sachets forwards and remove from the
workstation.