Bugbox
UM-16 R5.0
Page 43 of 48
Bugbox User Manual
Workstation malfunction
In the event of a workstation malfunction, please check section 7.6 for a list of common problems and
solutions. If you cannot find a solution to your problem, please contact your local distributor, quoting
the serial number of your workstation.
Common problems and solutions
Workstation general problems
Please consult the list shown below as a reference in the event of a malfunction of your workstation.
Figure 33: Common problems and solutions
Problem
Cause
Solution
The workstation will not
switch on
The workstation is not
plugged in
The mains socket is not
switched on
The mains plug fuse has
blown
The mains fuse has
blown
Plug the workstation into the mains
Switch the mains power on
UK users: Replace the mains plug fuse
(see section 7.2.4, for more details)
Replace the mains fuses. See section
7.2.5 for more details.
The workstation
temperature will not go
above ambient plus 5
o
C
The heater mat thermal
protection switch has
activated
The heater mat fuse has
blown
Wait for the heater mat thermal
protection switch to reset.
Contact your local distributor
The LED indicators are not
illuminating
24V power supply fuse
has blown
Contact your local distributor
Devices plugged into the
internal socket (if fitted) are
not powering up
The power switch on the
device is not switched
on
The plug for the device is
not fully inserted into
the socket
The fuse in the device
has blown and/ or the
device has failed
The internal socket fuse
has blown
Turn the power switch on the device on
Check that the plug is correctly inserted
into the socket
Plug the device into another socket. If
the device fails to power up, the fuse
has blown and/ or the device has failed.
Refer to the owner’s manual for the
device.
Contact your local distributor
The internal light does not
illuminate when switched on
The bulb has blown
The fuse has blown
Contact your local distributor to
arrange replacement
The spot light does not
illuminate when switched on
The bulb has blown
The fuse has blown
Contact your local distributor to
arrange replacement
Plates are drying out
Too little humidity inside
workstation
Place a tray of water inside the
workstation as a water source.