Bugbox
UM-16 R5.0
Page 39 of 48
Bugbox User Manual
Remove the new sachets from their packaging.
Whilst lifting the floor tray at the front, slide the sachets into the sachet holder.
Lower the floor tray, ensuring that the floor tray leg has located in the floor tray locator (item
3 in Figure 27).
Close the glove ports using the glove port covers.
Reinstall the Ezee Sleeves (see section 7.2.2 for more details).
Reconnect the Bugbox to the mains power supply and switch the Bugbox on.
Allow the Bugbox to reach an anaerobic environment in the workstation chamber by leaving
the Bugbox for a period of approximately 1 hour.
Check that the environment in the workstation is anoxic using the anaerobic indicator strips
(see section 5.2 for more details).
If the environment in the workstation chamber is anaerobic, samples/ cultures may be
reintroduced to the workstation chamber via the interlock. See section 6.1.4 for more details.
Replacing the mains plug fuse – UK users only
To replace the mains plug fuse;
Switch off Bugbox and disconnect from the mains power supply.
Remove the plug from the mains socket.
Using a small flat bladed screw driver, remove the fuse cover from the mains plug. Figure 28
shows the fuse removal;
Figure 28: Mains plug fuse removal
Replace the fuse with a BS 1362 13A fuse.
Replace the plug in the mains socket.
Reconnect Bugbox to the mains power supply and switch on the Bugbox.