Safety equipment warnings and CE compliance statements 
 

2/24/2010 | 71-018 V11 EN  
D12R, D6R, D3R Users Manual 

 

in water or other cleaners or solvents. To clean the screen, take a soft water dampened cloth 
and 

gently

 wipe the surface.   

 
Technical assistance / Authorized Service Centers 

See our website at 

information. This information will be marked with an asterisk.  
 

 

 

 

 

Accessory interconnection and use 

The D3R/D6R/D12R are equipped with a footswitch standard. Other accessories, such as the 
AT101ZTX and safety lights are optional. Please see details about these accessories in later 
chapters in this manual. 

 
Intermittent operation limits 

At this time there are no intermittent operation limits to the use of the AWA unit. 
 

Installation requirements 

The unit may be operated 
 

1)

 

Flat on the bottom of the unit, 

2)

 

Flat on the back of the unit, or 

3)

 

Held at an angle using the rotating handle. 

 

There are no ventilation requirements. 
The unit is intended for use in Installation Category II (Portable Equipment) areas and 
pollution Degree II Environments where occasional non-conducting condensing pollution can 
be encountered. 

 
Unpacking the unit 

Carefully remove the following items from the shipping boxes. 
D3R, D6R or D12R 
Power cord 
Operation manual 
 

Pollution Degree II 

(From IEC 61010-1 3.6.6.2) Only non-conductive pollution occurs. However, temporary 
conductivity caused by condensation is expected. 

 
Power requirements 

Using the provided AC power cord, connect the unit to a grounded AC power source.  The 
unit’s power requirements are 100-240VAC, 50-60 Hz, 2 amps AC maximum current draw. 
The unit is fused using 2.5A fast blow fuses. Replace fuses with like type and rating. 
 

Environmental conditions 

1)

 

The unit has been tested for use up to 2000 m. 

2)

 

The tester should only be operated in temperatures ranging from 41 to 104 
degrees Fahrenheit (5° C to 40° C). 

3)

 

This unit is for use at a maximum relative humidity of 80% for temperatures up to 
31 °C decreasing linearly to 50% relative humidity at 40°C. This unit is intended for 
Installation Category II in a Pollution Degree II environment. 

 

 
 

Summary of Contents for D12R

Page 1: ...rby equipment Information furnished in this manual by Baker Instrument Company an SKF Group Company is believed to be accurate and reliable However Baker Instrument Company an SKF Group Company assume...

Page 2: ...Buyer returns Products accessories or options to the Baker SKF factory for warranty service Baker SKF will bear the cost of return packaging and shipping and insurance on the return shipment For Produ...

Page 3: ...ctions 13 On line labels 14 Safety precautions for setup 14 Initial tester power up and checkout 15 Using the footswitch 15 2 17 Test sequence voltages applicable standards 17 Recommended testing sequ...

Page 4: ...tion Notes 44 Single phase motors and two terminal devices 44 Determination of a fault 44 Form coils 45 Determination of a fault 45 Notes and tips for form coils 45 Three phase motors 46 Determination...

Page 5: ...nnection 69 Single phase lead connections 69 Note regarding 220 240 VAC units 69 Lifting the instrument 70 Supplement 2 71 PP185 power pack 71 Important safety information 71 Other Important Safety wa...

Page 6: ...Bar to bar armataure testing accessory 95 Appendix A Winding Faults 97 Appendix B Troubleshooting 99 Self help and diagnostics 99 Step 1 Basic information 99 Step 2 Applications or service problem 99...

Page 7: ...ury NOTICE NOTICE is the preferred signal word to address practices not related to personal injury Other Important Safety warnings Failure to follow these precautions could result in severe electrical...

Page 8: ...pecifications 8 Ground the product This product is grounded through the grounding conductor of the power cord To avoid electrical shock plug the power cord into a properly wired grounded receptacle be...

Page 9: ...tion requirements The unit is intended for use in Installation Category II Portable Equipment areas and pollution Degree II Environments where occasional non conducting condensing pollution can be enc...

Page 10: ...s amended by Council Directive 93 68 EC and Council Directive 2004 108 EC on the approximation of the laws related to Member States relating to the electromagnetic compatibility as amended by Council...

Page 11: ...Ray Tube CRT is the location where the tester displays test information At the Top menus corresponding to the four function keys above the CRT are shown The main portion of the screen displays the wa...

Page 12: ...nterlock is disabled when this function is selected and the Footswitch is depressed AUX The auxiliary position is for use with a 30kV Power Pack The display will also show wave patterns from accessory...

Page 13: ...rn 15 Leads energized indicator light This indicator light will illuminate when voltage is applied during a test V fast up When pressed this control will increase the applied voltage to a motor at an...

Page 14: ...4 Seconds Div or micro Amps Div Setting Seconds division is displayed here during Surge Testing The micro amps division setting is displayed here during HiPot Testing Safety precautions for setup Ther...

Page 15: ...y upon power up on and assures a positive ground is available to the tester If the instrument is not properly grounded the Open ground indicator will light and testing cannot proceed Check the supply...

Page 16: ...Instrument Overview 2 24 2010 71 018 V11 EN D12R D6R D3R Users Manual 16...

Page 17: ...ually low megohm value when compared to previous measurements or industry accepted limits for the type of insulation in the motor If a low megohm value is measured the motor should be inspected for gr...

Page 18: ...vation for even doing the DA test is to reduce the test time to 3 minutes instead of 10 minutes for the PI test when the PI test may not be worthwhile To date there are no accepted values for the DA H...

Page 19: ...1372 1029 575 469 1643 1232 600 490 1715 1286 2300 1878 6573 4930 4160 3397 11888 8916 6900 5634 19718 14789 13800 11268 39437 29578 IEC 34 15 V Line Per Unit 1 2 x 50 4E 5000 0 2us 65 480 392 6920 4...

Page 20: ...chines IEEE 429 1972 Evaluation of Sealed Insulation Systems for AC Electric Machinery Employing Form Wound Stator Coils IEEE 432 1992 Guide for Insulation Maintenance for Rotating Electrical Machiner...

Reviews: