oxy.IQ User’s Manual
1
Chapter 1. Features and Capabilities
Chapter 1. Features and Capabilities
1.1
Introduction
The
oxy.IQ
Panametrics Oxygen Transmitter
(see Figure 1 below) is a highly reliable and cost-effective two-wire,
loop-powered transmitter with a linearized 4 to 20 mA output. It measures oxygen content in ten ppm ranges
(10, 20, 50, 100, 200, 500, 1000, 2000, 5000 and 10000 ppm) and eight percentage ranges (1, 2, 5, 10, 21, 25, 50 and 100%).
All ranges are user-selectable. This compact transmitter uses proven sensor technology to accurately measure O
2
in
a variety of gases, even in hazardous (classified) locations.
Figure 1: oxy.IQ
1.2
Hazardous Location Certifications
When equipped with an optional Zener barrier or galvanic isolator, the oxy.IQ can be mounted in a hazardous
(classified) location. The oxy.IQ with intrinsically safe option is certified to USA, Canadian, ATEX, and international
IECEx IS requirements. The standard oxy.IQ is certified to USA, Canadian, EU ATEX and International IECEx div2/Zone 2
non-incendive requirements.
1.3
Applications
Some typical applications for the
oxy.IQ
Panametrics oxygen transmitter include the following:
• Glove box purge and leak detection
• Natural gas
• Semiconductor wafer machines
• Coating process machines
• Membrane air separators
• Inert welding gases
• Pure gaseous hydrocarbon streams
• Process monitoring of gaseous monomers
• Heat treating and bright annealing
Summary of Contents for Panametrics oxy.IQ
Page 1: ...panametrics com BH024C11 EN G oxy IQ Panametrics Oxygen Transmitter User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Chapter 1 Features and Capabilities 4 oxy IQ User s Manual no content intended for this page...
Page 26: ...Chapter 5 The Service Menu 18 oxy IQ User s Manual no content intended for this page...
Page 30: ...Chapter 6 Specifications 22 oxy IQ User s Manual no content intended for this page...
Page 42: ...Appendix C Order String 34 oxy IQ User s Manual no content intended for this page...
Page 50: ...Appendix E Cell Models 42 oxy IQ User s Manual no content intended for this page...
Page 54: ...Warranty 48 oxy IQ User s Manual no content intended for this page...
Page 55: ......