620G Instruction Manual | 29
Error Indications
3.1.2 Calibrator Screen Information
Figure
3
-
3: Calibrator Screen Information Example
3.2 Error Indications
1. Make sure that the range is correct.
2. Make sure that all the related equipment and connections are serviceable.
1 Function Symbol
2 Channel Direction
3 Channel Function
4 Sensor Type
5 Utility Icon
6 Measurement Reading
7 Measurement Unit
8 Channel Settings
9 Hold/Pause Readings
10 Home or Dashboard
11 Back
12 Additional Measurement Data
13 Sensor Full-scale
14 Process Icon
Display
Condition
<<<<<
Under Range: The display shows this symbol for this condition:
Reading < 110% Negative Full Scale (Pressure)
Reading < 102% Negative Full Scale (Electrical)
>>>>>
Over Range: The display shows this symbol for this condition:
Reading > 110% Positive Full Scale (Pressure)
Reading > 102% Positive Full Scale (Electrical)