Recommended protective distances between portable and mobile HF communication devices and the model BaehrTec 

S1200  

The model BaehrTec S1200 is intended for use in an ELECTROMAGNETIC ENVIRONMENT, in which the emitted HF-
disturbances are controlled. The customer or the user of the model BaehrTec S1200 can help prevent, electromagnetic 
disturbances by observing the minimum distances between portable and mobile HF communication facilities (transmitters) and 
the model BaehrTec S1200, as given below according to the maximum output of the communication facilities  

 

Nominal output 

of the transmitter

 

 

Protective distance according to transmitter frequency 

150 kHz to 80 MHz 

 

= 1.2

P

 

80 MHz to 800 MHz 

 

= 1.2

P

 

800 MHz to 2.5 GHz 

 

= 2.3

P

 

0.01 

0.12 

0.12 

0.23 

0.1 

0.38 

0.38 

0.73 

1.2 

1.2 

2.3 

10 

3.8 

3.8 

7.3 

100 

12 

12 

23 

The distance can be determined for transmitters whose nominal output is not given in the table above using the formula that 
belongs to the respective column, where 

P

  is the nominal output of the transmitter in Watt (W) according to the specification 

provided by the manufacturer. 
 
NOTE 1 An additional factor of 10/3 was used to calculate the recommended protective distance from transmitters in the 
frequency range 80 MHz to 2.5 GHz in order to lower the probability of a disturbance being caused by a mobile/portable device 
unintentionally brought into the PATIENT area. 
 
NOTE 2 These guidelines may not apply in all situations. The distribution of electromagnetic waves is affected by absorption and 
reflections from buildings, objects and people.  

Spare parts/accessories 

 

Baehr cloths 

Art-No.: 11000

 

 

Baehr cloths refill bag 

Art-No.: 11001

 

 

Baehr quick surface disinfectant 

Art-No.: 11015

 

 

Baehr alcohol 

Art-No.: 11032

 

 

Brushes 

Art-No.: 34916

 

 

Power cable case, short for BaehrTec A2000/S1200 

Art-No.: 40796

 

 

Power cable with straight connector 

Art-No.: 20970006

 

 

Fine filter S1200 

Article-No.: 22000001 

 

Baehr spray solution for Baehr spray technology devices

 

Art-No.: 22333 

 
 
 
Contact address & manufacturer 

 

Gustav Baehr GmbH 
Max-Eyth-Strasse 39 
D-71332 Waiblingen 
Tel.: 07151 95 90 20 
Fax: +49 7151   1 84 44 
www.baehrshop.de 
 
 

 

Summary of Contents for BaehrTec S1200

Page 1: ...Operating Manual BaehrTec S1200 English V100 Status 09 2015 Page 1 48 BaehrTec S1200 Spray pedicure device Handpiece included Operating Manual...

Page 2: ...nd handpiece holder 16 4 1 3 Rear view 17 4 1 4 Bottom view 18 4 1 5 Top view with closed tank lid 19 4 1 6 Top view open tank lid 20 4 1 7 Tank lid inside 21 4 2 Description handpiece 22 4 3 Descript...

Page 3: ...arts or the like If the device falls this can cause severe damage and injuries To guarantee that the device functions reliably protect it from cold and also from overheating Therefore avoid temperatur...

Page 4: ...nd or if the device housing is damaged In the event of lighting strike and or overvoltage the device may be damaged For this reason we also recommend installing overvoltage protection and removing the...

Page 5: ...with the number of the certification authority A product bearing this label fulfils the requirements of the corresponding EU guideline the applicable European Standard 2 1 2 Type label with output det...

Page 6: ...operated The following applies to the device Operating time 15 minutes maximum pause time 10 min minimum These approved operating times correspond to the common work procedure in podology pedicure Fu...

Page 7: ...or the instrument currently in use is not exceeded In addition to the Easy Speed concept the BaehrTec S1200 has a colour display that provides you with further information and which will facilitate ad...

Page 8: ...allosities Other types and fields of application are carried out at your own risk and may conceal dangers No form of misappropriation is permitted Improper use may lead to damage to persons or objects...

Page 9: ...any other loose objects such as wipes cotton wool or the like can enter the area of the rotating tools A hair net must be worn where necessary The operating staff must bear in mind that the particles...

Page 10: ...must be operated according to the requirements of these operating instructions portable and mobile high frequency communication facilities may affect the functionality of electrical devices in order...

Page 11: ...e g pacemakers Please ask your patients before beginning the treatment whether they have such a system Due to the complex interactions between electric devices and mobile telephones it is possible th...

Page 12: ...e utmost care It contains important information on protecting yourself others and the device from damage The description used in this section BaehrTec S1200 refers both to the control device and the h...

Page 13: ...Please send the device to customer service with an error description Avoid significant differences in temperature This can cause dampness condensation Protect your BaehrTec S1200 from frost The devic...

Page 14: ...risk of injury As shown in the illustration you should avoid pulling motions during your work as this may cause the instruments to slip out Please take care not to exert too much pressure when using...

Page 15: ...LEDs 5a 12a 13 Push button to switch handpiece motor on and off Push buttons handpiece speed 14 und 15 This allows you to set the speed in the range of 6 000 20 000 rpm in increments of 1 000 and in t...

Page 16: ...Manual BaehrTec S1200 Page 16 48 English V100 Status 09 2015 4 1 2 Side view with handpiece outlet and handpiece holder Fig 5 29 Handpiece outlet 30 Protection from bending 31 Handpiece hose 32 Handpi...

Page 17: ...oving the storage tray from the case in order that the heat created by the device can escape from the case and to avoid potential head build up 33 Non heating device connector Only connect the connect...

Page 18: ...Bottom view Fig 7 37 Housing screws Must only be unfastened by the manufacturer 38 Ventilation holes Please ensure that these ventilation holes are never covered 39 Housing seal As soon as the housing...

Page 19: ...Operating Manual BaehrTec S1200 English V100 Status 09 2015 Page 19 48 4 1 5 Top view with closed tank lid Fig 8 41 Tank lid 42 Membrane lid...

Page 20: ...Operating Manual BaehrTec S1200 Page 20 48 English V100 Status 09 2015 4 1 6 Top view open tank lid Fig 9 43 Fine filter 44 Filing level contacts...

Page 21: ...Operating Manual BaehrTec S1200 English V100 Status 09 2015 Page 21 48 4 1 7 Tank lid inside Fig 10 45 Ventilation holes for ventilation membrane 46 O ring for tank lid 47 Screws...

Page 22: ...tus 09 2015 4 2 Description handpiece Fig 11 30 Protection from bending 31 Handpiece feed line 48 Handpiece button 49 Handpiece housing 50 Handpiece Fig 12 51 Opening for DIN instrument with shank dia...

Page 23: ...hat this is never exceeded since it can result in serious injuries to the patient and the operator Vibrations can also occur at the handpiece Please keep your safety and the safety of your patient in...

Page 24: ...y It is possible to work with the handpiece motor switched on when the spray function is deactivated However it is not possible to work with an activated spray function is the handpiece motor is not r...

Page 25: ...the spray function is not currently active at the selected spray level spray off This symbol is shown when the spray function is currently active at the selected spray level spray on Can only appear w...

Page 26: ...otor instrument rotation direction The instrument rotation direction is shown in the display Direction of rotation right Direction of rotation left The handpiece motor is running when the arrow is sho...

Page 27: ...automatically switched on hardly noticeable This was a conscious design decision since this prevents the handpiece from getting warm quickly Service control The BaehrTec S1200 has a servicing interval...

Page 28: ...iption Language German English French Italian Japanese Russian Spanish Set the desired language here Sounds Different sounds sounds off Here you can select the sound that will sound when you press a b...

Page 29: ...tly viewing The settings that have just been made will be indicated to you by the x on the right hand edge of the display 4 6 Home Function button 17 This button has two functions By briefly pressing...

Page 30: ...cannot be guaranteed that your instrument can be held firmly in the handpiece Caution risk of injury Never attempt to insert or remove instruments into out of the opening 51 when the handpiece motor i...

Page 31: ...00 rpm button 6 by presing the buttons 5 12 Now the LED for 10 000 rpm lights up automatically 6a This reminds you that a different maximum instrument diameter must be used The instrument heads indica...

Page 32: ...andpiece motor is running 13 You can make changes o before the handpiece motor is running the device runs using the factory settings when the handpiece motor is switched on o while the handpiece motor...

Page 33: ...ere may be a slight lightening or dulling of surfaces However this will have no effect on the function or the safety of the device Never immerse the device in water or any other liquids as this poses...

Page 34: ...the tank empty it completely immediately and rinse it with clear tap water Use a microfibre cloth to clean the tank wall if they are dirty Should the spray function malfunction at some point despite o...

Page 35: ...th your handpiece to stop even though the motor is running This can be accelerated even further by creams oils fats or the like The consequence The clamping mechanism is no longer functional or only f...

Page 36: ...he devicce while the handpiece motor is still running Risk of injury In addition the clamping mechanism will suffer damage 2 Put on a disposable glove or pull a finger cot over the thumb on the hand w...

Page 37: ...ed video instructions on cleaning the Easy Clean grips to supplement this information sheet please go to www fusspflege com easy clean To access the website directly please scan the QR code with your...

Page 38: ...ays in a perfect hygienic condition before use For hygiene reasons please clean the instrument changing aid after use under running water and then with Baehr quick surface disinfectant Art No 11015 Th...

Page 39: ...the instrument changing aid so that the instrument does not drop when removing Fig 24 Carefully remove the instrument from the clamping mechanism on the handpiece with the instrument changing aid Fig...

Page 40: ...ely with disinfecting agent Set pedicure device to spray level 6 and let it run for at least 45 seconds after the first spray jet Allow the disinfectant to work observe soaking time e g 7 5 solution 5...

Page 41: ...device which is caused by improper handling or falling is not covered by the guarantee There is no liability for defects and their consequences which are the result of improper intervention or modific...

Page 42: ...power supply before checking the fuses Fault Cause Solution None Spray ignored filling level indicator Fine filter in tank clogged see section Care Filling level contacts soiled see section Care Regul...

Page 43: ...nstrument Work with small instruments until larger instruments start to clamp securely again During operation always make sure that the handpiece tip and the instruments are clean Reduce pressure Avoi...

Page 44: ...p H Switch off your device at the main switch and disconnect the device from the mains Use a screwdriver to open the fuse compartment on the device socket Fig 26 Replace the microfuses in the compartm...

Page 45: ...ng voltage 28 V Controller dimensions W x D x H in mm without handpiece outlet without handpiece holder and without filler cap Approx 220 x 195 x 110 Weight inkl Handpiece without power cable Approx 2...

Page 46: ...Contact discharge 8 kV Air discharge Floorings should be made of wood or concrete or covered with tiles If the floor is covered with synthetic material the relative humidity must be at least 30 Electr...

Page 47: ...radios is lower than the CONFORMANCE LEVEL b at all frequencies according to a test conducted on location a Interference is possible in the vicinity of devices marked with the following symbols NOTE 1...

Page 48: ...sing the formula that belongs to the respective column where P is the nominal output of the transmitter in Watt W according to the specification provided by the manufacturer NOTE 1 An additional facto...

Reviews: