Checksum & serial number 

Page 10

 

LM_OG-AM_BA_02_1163

 

May 2019 

11.

 

CHECKSUM & SERIAL NUMBER 

Pressing 

RESET

 for 3 seconds will display the checksum and serial number in the bottom line. 

 

 

 

 

 

 

12.

 

STATUS MESSAGES 

 

 

 

 

 

 

 

 

 

 

 

13.

 

CHANGING THE BATTERY 

Important – Before starting the replacement: 

The battery has to be replaced within 30 minutes, otherwise you will lose the date and time setting. The display would show 

"Date" in the bottom which indicates that the date is lost. The quantity display will change to dashes "- - - -". To reconfigure 

the date, you would have to remove the register, which is sealed by a MID label. 

Please proceed as follows: 

 

 

 

 

Picture 1: 

Loosen the battery cover 

Picture 2: 

Take out the battery 

Picture 3: 

Insert the new battery and 

press the reset button to wake 

up the register 

Picture 4: 

Insert the battery cover, then 

screw the battery cover tight 

Battery: Lithium CR123A 

 

 

  3A33.07 

L

 

 

AC001 

TOTAL

 

3A33  

= Checksum over the MID relevant source code 

07  

= Sequential version number, not MID relevant 

Checksum 

= The year of production 

AC001  = Sequential alphanumeric number 

Serial number 

 

0.000 

L

 

 

Floir 

TOTAL

 

Wrong flow direction. Actual flow does not match with the programmed flow direction. FLDIR 
flashes in the bottom line, the quantity will change to dashes. See 

chapter 6 - step 8

: Programming of 

LM OG-AM and LM OG-TAERM series. 

 

-.--- 

L

 

 

Reed 

TOTAL

 

Channel missing due to broken reed switch REED flashes in the bottom line, the quantity changes 

to dashes. 

Call service for replacement. 

 

-.--- 

L

 

 

Datum 

TOTAL

 

Lost date due to power failure, i.e. slow battery exchange. 

DATE flashes in the bottom line, meter will not count anymore until the date has been entered 

again. Quantity changes to dashes. See 

chapter 13

Changing the battery 

Summary of Contents for Oval Gear LM OG-AM

Page 1: ...LM OG AM series Oval gear meters with MID approved register and integrated datalogger LM_OG AM_BA_02_1163 May 2019 User Manual LM OG AM LM OG TAERM LM OG CNDAM...

Page 2: ...conform meters 5 5 8 Installation procedure LM OG 5 6 Programming of LM OG AM and LM OG TAERM series 6 7 Calculating the correction factor 7 7 1 Calculation based default settings 7 7 2 Recalculation...

Page 3: ...table surface that may allow it to fall Never place the units above a radiator or heating unit Route all cabling away from potential hazards Isolate from the mains before removing any covers Power con...

Page 4: ...ation match with those of the lube meter for example connections pressure flow range and medium Once the meter has been installed please make sure that no air pressure shocks or particles can damage t...

Page 5: ...technicians or the officials of the national authorities MID 4 WARRANTY Badger Meter warrants meters and parts manufactured by it and supplied hereunder to be free from defects in materials and workma...

Page 6: ...front of the register No data are lost they are securely saved in an EEPROM memory except of the date and time To keep the date and time settings you have to reinsert the battery within 30 minutes See...

Page 7: ...dispense system or by an insufficiently areated oil pipe pressure shocks This can be prevented by installing an appropriate non return valve FLDIR will indicate a reverse flow quantity display will ch...

Page 8: ...arrow for the flow direction flashes now Use RESET to change or TOTAL to move forward 03 The measuring unit will flash and can be changed by pressing the RESET button L GAL QT PT To move forward pres...

Page 9: ...ctor would be for example 0 9960 Calculation 0 9960 x 1 003 0 9989 The new correction factor is 0 9989 8 PULSE OUTPUT AT REGISTER OG TAERM The pulse output on the LM OG TAERM register has a rating of...

Page 10: ...or more than 2s will bring you to the next lower menu level Pressing the TOTAL button once will bring you to the next newer data entry no matter if you are on the Date Time or Quantity menu level If y...

Page 11: ...e register will automatically move back to the standard display without saving the last entry 10 DISPLAY CORRECTION FACTOR To verify the programmed correction factor press TOTAL and RESET for 2 second...

Page 12: ...wake up the register Picture 4 Insert the battery cover then screw the battery cover tight Battery Lithium CR123A 3A33 07 L AC001 TOTAL 3A33 Checksum over the MID relevant source code 07 Sequential ve...

Page 13: ...ll CE declarations applying to this meter series under www badgermeter de en downloads ce declarations 16 DIN ISO AND EMC CERTIFICATE Please refer to our valid DIN ISO and EMC certificate under www ba...

Page 14: ...Return of goods for repair Harmlessness declaration Page 12 LM_OG AM_BA_02_1163 May 2019...

Page 15: ...Return of goods for repair Harmlessness declaration May 2019 LM_OG AM_BA_02_1163 Page 13...

Page 16: ...www badgermeter de...

Reviews: