Specifications
June 2021
SGN-UM-00282-EN-05
Page 12
SPECIFICATIONS
External Power
Input Voltage
8 5…30V DC (diode protected)
Maximum Current
Draw
18 mA (using internal resistor @ 30V DC input)
Environmental
Operating
Temperature
–22…158° F (–30…70° C)
Altitude
2000 m
Use
Indoor/Outdoor
Humidity
0…90% non-condensing
Inputs (Magnetic
Pickup)
Frequency Range 0…4000 Hz
Trigger Sensitivity 30 mV p-p…30V p-p
Output Signal
Max Voltage
30V DC
Max Power
0 25 W
Pulse Output
(using internal
pullup resistor)
Maximum Current 8 mA
VH =
Power input voltage – 0 7V DC
VL =
Less then 0 4V @ maximum input power
Internal Pullup
Resistor
3 6 kΩ (enabled/disabled by jumper)
Pulse Output
(using external
pullup resistor)
Maximum Current 100 mA
VH =
Input voltage to external pullup resistor
VL =
[VH /(selected resistor value + 47 Ω)] × 47 Ω
Pulse Length
150 µs, 1 ms, 25 ms, 100 ms, 500 ms, 1 s, or
auto mode
Enclosure
Killark aluminum capped elbow Y-3 Class I, Div 1 & 2, Groups C & D;
Class II, Div 1 & 2, Groups E, F and G; Class III
Agency Listings
CSA
Ordinary Locations
CAN/CSA C22 2 No 61010-1-12, UL Std
No 61010-1 (3rd Edition)
Pollution Degree 2
Overvoltage Category I
MPORTANT
I
For this CSA rating to be valid, the circuit board must be mounted in a certified Killark
one-inch model Y-3 conduit outlet box.