Badger Meter Blancett B220-885 User Manual Download Page 10

Startup

June 2021

SGN-UM-00282-EN-05

Page 10 

Status: Com1

9/23/2008

File Options Tools

Program Values

Board Values

Program

Read

Verify

Status:   Idle

Com Port 1

Pulse Output

High
Low

K-Factor

Pulse Width

150us

Auto

1s

500ms

100ms

25ms

1ms

Pulse Output

High
Low

K-Factor

Pulse Width

150us

Auto

1s

500ms

100ms

25ms

1ms

File

Tools Version

Program
Read
Verify
Com Port

Blancett K-Factor Programming Software

Status: Com1

9/23/2008

File Options Tools

Program Values

Board Values

Program

Read

Verify

Status:   Idle

Com Port 1

Pulse Output

High
Low

K-Factor

Pulse Width

150us

Auto

1s

500ms

100ms

25ms

1ms

Pulse Output

High
Low

K-Factor

Pulse Width

150us

Auto

1s

500ms

100ms

25ms

1ms

File

Tools Version

Blancett K-Factor Programming Software

Com 1
Com 2
Com 3
Com 4
Com 5
Com 6
Com 7
Com 8
Com 9
Com 10
Com 11
Com 12
Com 13
Com 14
Com 15
Com 16

Program
Read
Verify

Com Port

Figure 7:  Tools drop down

Figure 8:  Com port menu drop down

If the Com port selected is invalid, the software shows the message ERROR – 

Invalid Com Port when trying to program the board  If the incorrect Com port is 

selected, or if there is a problem with the cable, the software shows the message  

"No Response" after trying to program the board 

Setting the K-Factor

The K-factor is the ratio of input pulses per each output pulse, for example, if the 

K-factor is set to 1, then each input pulse would yield one output pulse  The range 

that the K-factor can be set to is 1…999,999,999  The K-factor is set by entering it 

in the Program Values column of the software in the K-factor field  

Setting the Output Pulse Width

The output pulse width is the length of time the pulse remains active before 

resetting to its original state  The K-factor scaler has a total of seven different pulse 

widths  The Auto option does not restrain the output pulse to a specific length  

Instead, it varies and is dependent on output frequency  The higher the output 

frequency, the shorter the pulse width output  The lower the frequency output, 

the longer the pulse width output  This option turns off the Pulse Output selection 

buttons because they do not apply in this mode  
Some end devices require the pulse to be a certain length or longer for proper 

detection of each incoming pulse  For these devices, select a pulse width that is 

long enough for the end device to recognize it  
The pulse width option is set by clicking the required pulse width radio button in 

the Program Values column of the software  

Setting the Output Level Normally High or Normally Low

Most end devices are unaffected by this setting, but the K-factor scaler can invert 

the output pulse level  This option is set by clicking the required pulse output 

radio button in the Program Values column of the software 

Summary of Contents for Blancett B220-885

Page 1: ...SGN UM 00282 EN 05 June 2021 User Manual K Factor Scaler B220 885 K Factor Scaler and B220 900 Programming Software Kit...

Page 2: ...afety 3 Terminology and Symbols 3 Considerations 3 Electrical Symbols 4 Introduction 4 Operating Principle 5 Installation 6 Electrical Connections 6 Enclosure Mounting Necessary for CSA Certification...

Page 3: ...f the purchaser SAFETY Terminology and Symbols Indicates a hazardous situation which if not avoided will result in death or serious personal injury WARNING Indicates a hazardous situation which if not...

Page 4: ...urbine meter to accumulate into a unit recognizable by an end device such as a PLC for counting and display The use of different K factor values allows the device to display in any number of volumetri...

Page 5: ...ugh the flow meter The K factor scaler input amplifier modifies the signal generated by the turbine The amplifier sends the modified signal to an onboard microcontroller which counts pulses up to a pr...

Page 6: ...r The turbine meter connections are non polarized Use shielded twisted pair wire for this connection to the turbine meter Pulse Output Either the internal or external pullup resistor must be used for...

Page 7: ...source This power source is the maximum input voltage of the pulse to the receiving device Use the equation below to determine the resistor value R Supply Voltage Current Where R Resistor value in hm...

Page 8: ...ERNAL 3 6K PULL UP RESISTOR JUMPER PROGRAMMING PORT Racine WI U S A K FACTOR SCALER INPUT 8 5 30 VDC MAX INPUT CURRENT 18 mA C US FILE 215035 B220 886 Flow Meters 1 2 3 4 5 6 1 2 3 4 5 6 7 8 9 10 Tab...

Page 9: ...t 1 Pulse Output High Low K Factor Pulse Width 150us Auto 1s 500ms 100ms 25ms 1ms Pulse Output High Low K Factor Pulse Width 150us Auto 1s 500ms 100ms 25ms 1ms File Tools Version Blancett K Factor Pro...

Page 10: ...t pulses per each output pulse for example if the K factor is set to 1 then each input pulse would yield one output pulse The range that the K factor can be set to is 1 999 999 999 The K factor is set...

Page 11: ...The K factor scaler retains downloaded values in the memory if it is disconnected from power Press Read to load the current information from the K factor scaler and display it in the Board Values colu...

Page 12: ...put voltage 0 7V DC VL Less then 0 4V maximum input power Internal Pullup Resistor 3 6 k enabled disabled by jumper Pulse Output using external pullup resistor Maximum Current 100 mA VH Input voltage...

Page 13: ...the number of counts 1000 by the number of seconds 60 1000 60 16 6666 Hz If you were looking at the pulse output on a frequency counter an output frequency of 16 666 Hz would be equal to 1 gpm If the...

Page 14: ...nd the inside pipe diameter Also keep in mind that one US gallon of liquid is equal to 231 cubic inches Example 3 Known values are Velocity 4 3 ft sec Inside Diameter of Pipe 3 068 in Find the area of...

Page 15: ...ge 15 SGN UM 00282 EN 05 June 2021 Known values are Frequency 700 Hz By measurement Flow Rate 99 1 gpm By calculation 700 Hz x 60 sec 42 000 pulses per gallon K factor 42 000 pulses per min 99 1 gpm 4...

Page 16: ...er Meter Inc Other trademarks appearing in this document are the property of their respective entities Due to continuous research product improvements and enhancements Badger Meter reserves the right...