Instruction 24-9448
3-13
Initial Setup
3.13 Zoom-Display Selection
Combustion test data in the Run/Hold screen can
be shown with enlarged characters to make view-
ing easier. The operator can set zoom levels to
Standard, 2X, or 3X. The Standard zoom setting
will display 8 lines of combustion test data at one
time, 2X which will display 4 lines of combustion
test data with enlarged characters, and 3X which
will display 3 lines of combustion test data with en-
larged characters. The operator can scroll through
the complete list of measured and calculated data
no matter what zoom level has been selected.
Select desired zoom level as follows:
1. Display the MAIN MENU by pressing the
MENU (F2)
button. If necessary, press
ESC
until MENU appears above
F2
.
2. Use the
buttons to highlight SETUP, and
then press
ENT
to display the SETUP MENU.
3. Use the
buttons to highlight ZOOM, and
then press
ENT
to display the ZOOM screen.
4. Use the
buttons to select the desired Zoom
level. Options include STANDARD, 2X,
and 3X. Standard will display eight lines of
Combustion test data, 2X will display four,
and 3X will display 3.
5. Press
ENT
to save the selection and re-display
the SETUP MENU.
MAIN MENU
FUEL
PRESSURE
TEMPERATURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1
F2
F3
MENU
SETUP MENU
(more)
TIME
O
2
REFERENCE
PRINT PRESSURE
ZOOM
BATTERY CHARGER
LOGGING
(more)
MENU
F1
F2
F3
ZOOM
Standard
2x
3x
05/26/06 09:25:30 AM
MENU
F1
F2
F3
O
2
4.0
%
CO
12
ppm
EFF 82.6
%
PRINT MENU SAVE
F1
F2
F3
Hold NGas
Instruction 24-9448
5-9
Calibration
If the sensor’s output is too low to be usable, then the message “Bad
Calibration Sensor End of Life, Entry Not Saved” will appear. The sen-
sor will now be marked as being
BAD
in the DIAGNOSTICS screen.
NOTE:
H
2
calibration can be bypassed by pressing the
ESC
button, after which the CALIBRATION LIST screen is re-dis-
played. Skip to Step 10 if the
ESC
button was pressed.
6. Turn OFF the regulator of calibration fixture and remove the CO
cylinder.
7. Attach a combination 1,000 ppm CO and
1,000 ppm H
2
cylinder to the regula-
tor of the calibration fixture, and then
use the
and buttons to enter an
Applied value that exactly equals the CO
concentration stamped on the cylinder.
8. Adjust the regulator for a flowmeter indica-
tion of approximately 2 SCFH, and then
press
ENT
to display the CALIBRATE H
2
screen.
The calibration range is from 400 to 1,500 ppm. An attempt to cali-
brate outside this range will cause the message “Bad Calibration
Wrong CAL Entry” to appear in the following
step.
9. Wait until the Measured reading stabilizes
(approximately 3 minutes), and then press
ENT
to calibrate the H
2
Measured value to
that of the Applied value; after which the mes-
sage “Good Calibration” should briefly appear
followed by the CALIBRATION LIST screen
being re-displayed.
If the sensor’s output is low, but still usable,
then the message “Good Calibration WARNING Low Sensor”
will appear. The sensor will now be marked as being
Low
in the
DIAGNOSTICS screen.
If the sensor’s output is too low to be usable, then the message
“Bad Calibration Sensor End of Life” will appear followed by the
CALIBRATION LIST screen being re-displayed. The sensor will now
be marked as being
BAD
in the DIAGNOSTICS screen.
TEST GAS CO/H
2
Enter CO VALUE
1000 PPM
06/27/06 11:55:30 PM
F1
F2
F3
CALIBRATE H
2
Measured 1000 PPM
Applied 1000 PPM
Press ENT
to calibrate
06/27/06 11:55:30 PM
F1
F2
F3