CAUTION:
In order to comply with CE EN61326-1:2006 standards, all electrical
connections to the three terminal blocks on the back of the analyzer are to be made
using shielded cable conductors. Each electrical connection to the terminal blocks
must be looped through the ferrite clamp provided. If more than one shielded cable
conductor is used, the individual wires from the different cables may share the same ferrite clamp.
Strip the cable jacket back about 8 inches to ensure all the individual wires fit through the ferrite
clamp. Additional ferrite clamps may be purchased from Neutronics if necessary (See Appendix F:
Spare Parts).
CAUTION:
Also, in order to comply with CE EN61326-1:2006 standards, all shielded cable
conductors must be earth grounded at both ends of the cable. The shield drain wire from the cable
conductors must be connected to the earth ground screw terminal located at the back of the
analyzer. It is also recommended that the exposed drain wire be covered with shrink tubing and
the end be terminated with a #6 ring terminal before connecting it to the earth ground screw
terminal. The recommended shrink tubing size is 1
/8” expanded to 1/16” reduced. See Figure 5.