1100-2293
Revision 0
8
MGS-460 Quick Start Guide
1. Reinstall rubber gaskets in remote sensor and transmitter.
Remote Sensor:
Place inner skirt facing the main enclosure.
(If the rubber gasket
doesn’t fit, turn 90°.)
Transmitter:
Place the side with two grooves facing the main enclosure and
the edge with two bumps on the top.
2. Reconnect the ribbon cable from the sensor to the PCBA.
3. Ensure no cables are interferring with the sensor module and close the lid.
4. Using a 5/32″
(4 mm)
hex key / allen wrench, tighten the lid screws in an “X”
tightening pattern:
STEP 3 | Reinstall Sensor & Connect Lid
WARNING:
DO NOT leave excess cable inside of the gas detector housing.
Failure to comply may result in damage to the produc
t
.
IMPORTANT:
To achieve proper seal in the IP66 version, the lid screws
should be torqued to 15 to 20 lbf in (1.5 to 2.0 Nm.