1100-2295 Rev 1
37
MGS-408 User Manual
6.1.2 DISPLAY LAST FAULT
Displays historical fault code and listing. Intermittent fault conditions can be
reviewed here.
Main Menu
DIAGNOSTICS
LAST FAULT
press
OK
, to access review fault
conditions.
Use the
Up/Down arrows
to select a fault descriptor, then press
OK
for more
detailed information about the fault and possible remedies.
6.1.3 CLEAR FAULT
Resets the current active fault code to zero (no faults). Use to clear faults when see
FAULT LATCHING is enabled. See “Fault Latching” on page 18 for further information.
Main Menu
DIAGNOSTICS
CLEAR FAULT
press
OK
twice, to reset the active
fault code:
6.1.4 CLEAR LAST FAULT
Resets the historical last fault code to zero. Use to find recurring intermittent fault
conditions.
Main Menu
DIAGNOSTICS
CLEAR LAST FAULT
press
OK
twice, to reset the last
fault code.
6.1.5 CLEAR LAST SD FAULT
Resets the historical SD card fault code. Use to find recurring intermittent fault
conditions with the SD card or data logging.
Main Menu
DIAGNOSTICS
CLR LAST SD FAULT
press
OK
twice, to reset the
SD
fault
code.
6.1.6 POWER
Displays live power supply voltages and chassis temperature. Use when a SUPPLY
VOLTS ERR or CHASSIS TEMP HI is indicated as a fault.
Main Menu
DIAGNOSTICS
POWER
press
OK
, to view power supply voltage
information:
Voltage tolerances are as follows:
y
SUPPLY IN=24 V ±3
y
3.3V=3.3V ±0.3 V
y
SUPPLY OUT=24 V ±3
y
BAT= 1.8V to 3.6 V
(Coin cell voltage)
y
5V=5V ±0.3
y
T =60 to -20deg C
(Internal chassis temperature)