IEQ Chek
Operation Manual
1509-9001 Rev 1
25
Slot
Location Restrictions
1
O
2
(if used) MUST use this slot. Otherwise, it may contain a toxic gas
sensor that does not require bias voltage.
2
Toxic gas sensor that does not require bias voltage.
3
Toxic gas sensor requiring bias voltage (if used) MUST use this slot (e.g.,
NO). Otherwise, this slot may contain toxic gas sensor that does not
require bias voltage.
4, 5
CO
2
, PID, or catalytic gas sensor.
8. MAINTENANCE
Externally, the IEQ Chek
requires only cleaning with a damp cloth, wiping off the
exterior surface and basic inspection for obvious damage or problems. Internally,
sensors should be maintained because they have a specific life span. As they age,
they must be calibrated (zero and span adjusted) for accuracy. If one or more sensors
do not respond to span gas, check the age of the sensor as they may have expired.
Refer to the sensor replacement and calibration sections of this manual for more
details. Contact the factory for service, to replace expired sensors, or for any other
parts that are required.
BACHARACH, INC.
621 HUNT VALLEY CIRCLE
NEW KENSINGTON, PA 15068-7074
USA
TOLL FREE: 1-800-736-4666
FAX:
724-334-5001
E-MAIL:
WEBSITE:
www.MyBacharach.com
9. IEQ CHEK
TROUBLESHOOTING
IEQ Chek
does not communicate with IQ Chek
software, following the
recommendations listed below.
•
Verify the instrument is switched on and has completed its warm-up cycle.
•
Verify that the IEQ Chek
is in the normal display mode indicating all installed
sensors on the LCD.
•
If the above two functions do not rectify the problem, close IQ Chek
and reopen
it. Windows is sometimes slow to recognize certain programs.