Manual 1-Stage Mode
In the Manual 1-Stage mode the user starts a measurement at a high
flow rate and then waits for a stable leak rate reading to occur. During
this time the flow rate can be changed using the
Speed
function on the
Main Screen. After a stable leak rate reading is obtained, the user can
manually
save all final measured and calculated values in memory using
the
Save
function on the Expanded Main Screen. This mode is used
when more control of the leak rate measurement process is required.
Select Manual 1-Stage operation as follows:
1.
From the Expanded Main Screen, select
Menu
>
Operating Mode
.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
C a l i b r a t i o n
S e t F l o w U n i t s
M e n u M o d e
O p e r a t i n g M o d e
A c c e s s R e c o r d s
[ 0 0 0 1 ] ( M )
S - -
2.
Highlight
Manual 1-Stage
, and then press the
I/O
key.
0 2 / 2 1 / 0 3 0 9 : 4 5 : 3 0
A u t o m a t i c 2 - S t a g e
M a n u a l 2 - S t a g e
M a n u a l 1 - S t a g e
[ 0 0 0 5 ] ( M )
S - -
3.
Press
ESC
to return to the Main Screen. Observe that
(M)
should
appear at the bottom of the screen
Summary of Contents for HI FLOW Sampler
Page 8: ...Introduction Hi Flow Sampler Instruction 0055 9017 4 Notes...
Page 52: ...Calibration Hi Flow Sampler Instruction 0055 9017 48 Notes...
Page 62: ...Maintenance Hi Flow Sampler Instruction 0055 9017 58 Figure 5 9 Gas Flow System...
Page 68: ...Troubleshooting Hi Flow Sampler Instruction 0055 9017 64 Notes...
Page 71: ......