APPENDIX B. AGENCY APPROVALS
D
ECLARATION
OF
C
ONFORMITY
The manufacturer of the
products covered by this
declaration:
Bacharach, Inc.
621 Hunt Valley Circle
New Kensington, PA 15068
Year conformity is declared:
2010 (EN 14624), 2008 (all others)
Product(s):
Refrigerant Monitors
Model(s):
HGM-SZ and HGM-MZ (Single and Multi Zone Refrigerant Monitors) (EN 14624 only)
HGM-SZ, AGM-SZ, CO2-SZ, HGM-MZ, AGM-MZ, CO2-MZ and MZ-RD.
The undersigned hereby declares that the above referenced products are in conformity with the provisions of the following
standard(s) and is in accordance with the following directive(s).
Standard(s):
EN 14624
Performance of mobile leak detectors and of room controllers of halogenated refrigerants
EN 61326-1:2006 Electrical equipment for measurement, control, and laboratory use—EMC requirements: Part 1 General Requirements
EN 61010-1:2001 Safety requirement for control, measurement, and laboratory equipment
Directive(s):
2004/108/EC
EMC Directive
2006/95/EC
Low Voltage Directive
Signature:
Name:
Doug Keeports
Title:
VP of Product Development
Date:
06 December 2010
The technical documentation file required by this directive is maintained at the corporate headquarters of Bacharach, Inc.
European Standard EN14624: 2005 testing with R-134a (Halogen Gas Monitor Only).
Min. Sensitivity Threshold
1 ppm
Max. Sensitivity Threshold (within 1 ppm ± 10% reading accuracy range)
1,000 ppm
Max. Sensitivity Threshold within Instrument Reading Range
10,000 ppm
Zeroing Time from 1,000 ppm*
<12 seconds
Zeroing Time from 10,000 ppm*
<25 seconds
Reaction Time for detection of minimum threshold*
≤12 seconds
Min. Sensitivity threshold once max. threshold has been measured
1 ppm
Repeatability at 500 ppm
± 1% of reading
* Based on a minimum tubing length of one foot