Bacharach HGM-SZ Installation & Operation Manual Download Page 26

Single Zone Gas Monitor 

22 

P/N: 3015-4256 Rev 13 

SQUELCH

 – Sets a value of between 0.0 and 99.9 ppm that prevents 

the display of measurements below that value. Factory default is 0 ppm. 
For example, if the squelch setting is set to 50 ppm, then the monitor will 
not display measurements that are below that value. Use the 

Keypad

 to 

enter the desired value, and then press 

ENTER

  to save that value and 

return to the previous screen. 

 

ZONETEMP

  –  Used to enter the temperature of the area being 

monitored in °C, thus giving a  more accurate ppm reading. The factory 
default is 25°C. Use the 

Keypad

 to enter the desired temperature, and 

then press 

ENTER

 to save that value and return to the previous screen. 

 

CAL

 – Used to change the monitor’s calibration factor. This function is 

to be used ONLY with instructions from a Bacharach Service Center. 
Factory default is 1.000. 

 

P-CHK

  –  This 

Pressure Check 

function displays the current manifold 

pressure and the stored ambient pressure in psia, along with the 
difference between these two pressures and the current fault code. 

 

 

3.4.  Gas Alarms 

3.4.1.  Overview 

If the selected  gas  ppm  level  in the area being monitored exceeds  its  preset  Leak,  Spill,  or Evacuate 
Alarm level  (page  22),  the  SZ  unit  will detect this  alarm condition and turn ON  the  front panel 

ALARM

 

light. Additionally, an external alarm device may activate and the monitor’s internal audible alarm may 
sound if those features have been enabled (pages 14 and 22). 

3.4.2.  Clearing / Silencing a Gas Alarm 

Once the gas-alarm circuit has been triggered all alarm indicators remain turned ON, even after the 
detected gas level returns to normal. Press the 

SILENCE 

button to clear all gas-alarm indicators after the 

cause of the alarm has been cleared and the detected gas level has dropped below all alarm levels.  

Pressing the 

SILENCE 

button while a gas-alarm  condition  still  exists causes  the internal audible alarm 

and all external alarm devices that are connected to the relays to turn OFF for a period of time as set by 
the 

SILENCE 

function (page 22).  This will allow ventilation equipment tied to the SPILL and EVAC relays 

to continue operating. The front panel 

ALARM

 light remains ON, however, as an indication that an alarm 

condition still exists, but will automatically turn OFF once the gas level returns to normal. The alarm circuit 
will reactivate at the end of the silence period if the detected gas level is still above an alarm level. 

3.4.3.  Viewing the Gas Alarm Log 

From the 

Data Display

 screen, use the 

Keypad

 buttons to place the arrow (>) on the display next to the 

ALARMS

 function. Then press 

ENTER

 to display the alarm log. 

The alarm log shows  the type of alarm (LEAK, SPILL, or EVAC), plus the date and time it occurred.  If 
CLEAR is displayed, this indicates that an alarm was acknowledged at the date and time shown. 

Immediately after selecting the 

ALARM 

function, the most recent alarm event is displayed. In the example 

below, record  #03  shows that a Spill Alarm occurred on 11/10/10  at 15:06. Note that if more than 
30 alarm events have occurred, then the newest event overwrites the oldest.  The Alarm Log can be 
cleared as described on page 25. 

 

#03     SPILL @ 
11/10/10  15:06 

14.72 * AMB14.81 

 0.0

 

  

 

 R134A CALFACTOR 
     1.000 

TEMP AT ZONE = 
       25 degC 

SQUELCH BELOW 
     50.0ppm 

Summary of Contents for HGM-SZ

Page 1: ...tors HGM SZ Halogen and AGM SZ Ammonia Instruction 3015 4256 Installation Operation Maintenance Rev 13 January 2013 Patent 6 590 690 UL 61010 1 CSA 22 2 No 61010 1 EN 14624 Product Leadership Training...

Page 2: ...II ANY OBLIGA TION LIABILITY RIGHT CLAIM OR REMEDY IN CONTRACT OR TORT WHETHER OR NOT ARISING FROM BACHARACH S NEGLIGENCE ACTUAL OR IMPLIED The remedies of the Buyer shall be limited to those provide...

Page 3: ...ng Point 4 1 7 Specifications 5 SECTION 2 INSTALLATION 7 2 1 Installation Considerations 7 2 1 1 Warnings and Cautions 7 2 1 2 Inspection 7 2 1 3 Location of the Monitor 7 2 2 Mounting Instructions 7...

Page 4: ...3 9 2 Keypad Functions 26 3 9 3 First Diagnostic Screen 26 3 9 4 Second Diagnostic Screen 26 3 10 The Calibration Function 27 3 10 1 Overview 27 3 10 2 Calibration Procedure 27 3 10 3 Adjusting Calibr...

Page 5: ...zard Symbols Located on the Monitor This symbol indicates the need to consult this manual when opening the enclosure WARNING A potential risk exists if the operating instructions are not followed This...

Page 6: ...nt in the presence of flammable liquids vapors or aerosols Operation of any electrical equipment in such an environment constitutes a safety hazard 1 5 4 Proper Exhaust Venting WARNING It is imperativ...

Page 7: ...onitor retains a log of previous readings that can be easily accessed for analysis An audible alarm and front panel indicators are provided to signal alarm and fault conditions and relay contacts are...

Page 8: ...d settle below the leak point When the gas warms to room temperature it becomes lighter than air and tends to collect above the leak point Consequently the sampling point should be located as close as...

Page 9: ...R 227 R 23 R 401a MP39 R 402a HP80 R 402b HP81 R 408a R 409a R 500 R 503 HALON H1211 H1301 H2402 Other FA188 FC72 HFO1234YF N1230 R 420a R 424a R 426a R 427a R 438a Sensitivity All gasses 1 ppm Measur...

Page 10: ...cycles Sampling Mode Automatic continuous Re Zero Automatic approximately every 5 minutes System Noise Less than 50 dB A 10 feet 3m Monitoring Distance 500 ft 152 m maximum Conditioned Signal 4 20 mA...

Page 11: ...ice Center for assistance 2 1 3 Location of the Monitor Place the monitor in a location that will require no more than 500 ft 152 4 m of tubing total of sample line exhaust to reach the sampling area...

Page 12: ...securely against the mounting surface 2 3 Connecting Gas Sample and Exhaust Lines 2 3 1 Overview A single gas sample line must be run from the SZ unit to the area of the facility to be monitored An o...

Page 13: ...used should not exceed 500 ft 152 4 m The end of this line should be placed near the potential leak source and positioned to reduce the possibility of mists aerosols oil water dust or other contamina...

Page 14: ...rge line It may be mounted adjacent to the monitor Use upgrade kit 3015 5716 to convert an SZ with an internally mounted filter to an SZ with an externally mounted filter A line end filter P N 3015 34...

Page 15: ...one Halogen Gas Monitor NOTE The plastic cable tie surrounding the pump is for stability during shipping Please remove before operating If the unit gets shipped to Bacharach for service reinstall a pl...

Page 16: ...mmonia Gas Monitor NOTE The plastic cable tie surrounding the pump is to ensure safe handling during shipping Please remove before operating If the unit gets shipped to Bacharach for service reinstall...

Page 17: ...ing AC power ground wire green to the monitor s AC Input Ground Stud and then install a separate wire between the ground stud and the GND terminal A switch or circuit breaker rated 1 0 A 250 VAC with...

Page 18: ...LEAKLVL SPILLLVL and EVACLVL function screens page 22 2 7 2 Connection Use any of the service knockouts to gain cable access to the interior of the monitor Locate the External Alarm Relay Connector pa...

Page 19: ...Single Zone Gas Monitor P N 3015 4256 Rev 13 15 Figure 2 9 Typical AC External Alarm Relay 1 Wiring Figure 2 10 Typical DC External Alarm Relay 1 Wiring...

Page 20: ...remove the resistor or shorting wire from its terminals IMPORTANT The monitoring device must be isolated from ground floating NOTE A 100 Ohm W resistor must be connected to the 4 20 mA connector if no...

Page 21: ...one or more faults occur Alarm LED Red Flashes when detected gas level exceeds any of three preset alarm points Display Screen Displays current readings menus and parameters Enter Button Press this b...

Page 22: ...on has occurred page 14 Pressing the front panel SILENCE button will acknowledge an alarm and turn OFF all alarm indicators with the exception of the front panel ALARM light The alarm circuit will rea...

Page 23: ...process is performed on an as needed basis which is normally once every 6 to 8 minutes PRES CHK is displayed when the monitor is performing an atmospheric pressure check which is done to ensure the ac...

Page 24: ...ds of the last 30 fault events with the monitor s current fault status being shown when the Fault screen is first displayed After 30 events have been recorded the newest record overwrites the oldest E...

Page 25: ...ero and span checking note that when the LOOP function is selected the loop output is set to 20 mA and upon exiting this function the loop output is set to 4 mA AUDALRM Allows the monitor s internal a...

Page 26: ...triggered all alarm indicators remain turned ON even after the detected gas level returns to normal Press the SILENCE button to clear all gas alarm indicators after the cause of the alarm has been cl...

Page 27: ...play next to the FAULTS function Then press ENTER to display the fault log The fault log shows the monitor s current fault status If the fault is still present when the FAULTS function is selected the...

Page 28: ...the 4 20 mA connector page 16 0100 Zero Filter Fault There is gas present in the purge air Locate the monitor in an area containing fresh air or install a purge air line Refer to page 7 0200 Gain Set...

Page 29: ...oting a flow fault problem The monitor will trigger a flow fault if the pressure drop from ambient is less than 0 02 psi during a purge cycle and 0 06 psi during a measurement cycle 3 8 2 Keypad Funct...

Page 30: ...he current internal enclosure temperature in C Fault Code Current fault code A value of 0000 indicates that no faults are being detected Manifold Pressure Current manifold pressure in psia Purge Valve...

Page 31: ...d down to but not below 100 ppm A cylinder of a certified calibration gas must be used to ensure sampling occurs at ambient conditions A minimum sample size of 5 liters is required The SZ unit should...

Page 32: ...er which is mounted outside of the enclosure using the included bracket removes refrigerant gas from the purge air stream during the internal purging process External Filter Includes Mounting Bracket...

Page 33: ...nitor Install a new sample pump P N 3015 5176 4 7 End of Line Water Stop Optional The end of line water stops P N 3015 5512 prevent contaminates and moisture from entering the monitor This filter shou...

Page 34: ...0007 1655 Replacement Pump 3015 5176 Replacement Filter 0007 1656 Filters P N Mounting Bracket 0007 1657 Charcoal Zero Air HGM Only If external use 3015 3125 End of Line Water Stop Filter 3015 5512 I...

Page 35: ...Circle New Kensington PA 15068 7074 Phone 724 334 5000 800 736 4666 Fax 724 334 5723 Email help MyBacharach com Canada Bacharach of Canada Inc 20 Amber Street Unit 7 Markham Ontario L3R 5P4 Canada Ph...

Page 36: ...R404A 100 300 500 R407A 100 300 500 R407C 100 300 500 R409A 100 300 500 R410A 100 300 500 R500 100 300 500 R502 100 300 500 R503 100 300 500 R507 100 300 500 R508B 100 300 500 H1301 100 300 500 R408A...

Page 37: ...2 Response Time Flow Rate Gas Sample Line Feet Meters Response Time Seconds HGM Response Time Seconds AGM 0 0 0 100 30 5 20 40 200 61 40 60 300 91 4 56 90 400 123 80 120 500 152 4 105 180 Dependent o...

Page 38: ...easurement control and laboratory use EMC requirements Part 1 General Requirements EN 61010 1 2001 Safety requirement for control measurement and laboratory equipment Directive s 2004 108 EC EMC Direc...

Page 39: ...Single Zone Gas Monitor P N 3015 4256 Rev 13 35...

Page 40: ...ev 13 Headquarters 621 Hunt Valley Circle New Kensington PA 15068 7074 Toll Free 800 736 4666 Tel 1 724 334 5000 FAX 1 724 334 5001 Website www MyBacharach com E mail help MyBacharach com Printed in U...

Reviews: