Figure 2
At the setup screen you can view the current conditions by entering the diagnostic
screen “DIAG”. To make changes and adjustments you must first put the monitor in the
SERVICE MODE. Once in service mode select “SYSTEM”.
To access detector voltage and IR
voltage adjustment screen select
SERVICE MODE ENTRY – press
enter twice then select SYSTEM
and press enter.
To enter service mode:
From the main sampling screen (figure 1) press enter two times that will take you to the SETUP
screen (figure 2) then move the cursor to the Service Mode entry box, then press enter, it will now
say (service mode confirm), press enter again, it will now say, (service mode quit), press enter
again, it will now say (service confirm quit), press enter again and if you did it correctly it will say
(service mode entry) again.