Halogen Monitor
Operation
Instruction 3015-4584 Rev 9
11
•
TWA ALARM
– Audible alarm sounds when the TWA or twa, whichever is selected by the
2ndUNIT
Function
, exceeds the ALARMLVL value.
The audible alarm can be temporarily silenced by pressing the
SILENCE
button.
PPM LOG
– Contains records of the last 200 measurements. Each
record shows the measurement’s date, time, and ppm level.
Measurements are logged at an interval determined by the
LOG INT
function. After all 200 locations are filled the newest entries overwrite
the oldest entries.
Use the
Keypad
Up
and
Down
buttons to change the record number
by a factor of 1. Use the
Right
and
Left
buttons to change the record
number by a factor of 10. Press
ESC
to return to the previous screen.
To the right, record #025 shows that a gas level of 7 ppm was
measured on 07/11/2005 at 3:35 PM.
The PPM Log can be cleared using the information discussed in
Clearing the PPM, Alarms and Faults Data
on page 14.
LOG INT
– Sets the interval at which measurements are logged to
memory from 1 to 9999 minutes. Factory default is 1 minute. The
logged measurements can be viewed using the
PPM LOG
function.
Use the
Keypad
to enter the desired value, and then press
ENTER
to
save that value and return to the previous screen.
FAULTS
– Contains records of the last 30 fault events. The most
recent event is displayed when the
Fault
screen is first displayed. After
30 events have been recorded, the newest record overwrites the
oldest. Each record lists an event’s numeric fault code (refer to Section
2.11.4
Fault Codes
on page 14) plus the date and time at which the
event occurred. Use any of the
Keypad
buttons to scroll through the
other fault events. Press
ESC
to return to the previous screen.
To the right, record #15 shows that a Purge Flow Fault (fault code
<1000>) occurred on 07/11/2005 at 12:37 PM.
DIAG
– Enters the diagnostic function. Refer to Section 2.13.
SILENCE
– Sets the length of time the internal audible alarm is turned
OFF when the front panel
SILENCE
button is pressed. The factory default
is 30 seconds. If the cause of the gas-alarm has not been cleared at the
end of this time period, the internal audible alarm will reactivate. Use the
Keypad
to enter the desired time period, and then press
ENTER
to save
that value and return to the previous screen.
P-CHK
– The
Pressure Check
function displays the current manifold
pressure and the stored ambient pressure in psia, along with the
difference between these two pressures and the current fault code.
Refer to Section 2.15.
#025 7PPM @
07/11/05 15:35
14.07 * AMB14.05
-0.02dif <1000>
LOG INTERVAL IS
0001 min
4.2159v <0000>
29.05cD 14.58psi
SILENCE TIMEOUT
0030 sec
#15 <1000> @
07/11/05 12:37