background image

H-SEM 

Introduction 

Instruction 3015-4584 

 

3

 

1.8  Features and Capabilities 

 

Detects and measures CFC, HCFC, HFC and halogen gases 

 

Measures all gases up to 1,000 ppm with a sensitivity of 1 ppm 

 

Automatically logs up to 200 gas readings that can be later be recalled for analysis 

 

Eliminates false alarms with use of non-dispersive IR source and sample draw system 

 

Visual and audible gas alarm indictors that are turned ON when the detected gas level exceeds 
a user defined trip-point 

 

Extensive self diagnostics, providing both visual and audible indications when a fault occurs 

 

Battery powered, providing from 8–12 hours of operation 

 
 

1.9 Functional Overview 

1.9.1 General Description 

The Bacharach Halogen Safe Entry Monitor (hereafter referred to as the H-SEM unit) is a portable 

refrigerant gas monitor with a self contained purge-air supply. The H-SEM unit is designed to test an area 

for a specific refrigerant gas for the purpose of determining the presence and level of that gas in the area 
being tested. 
The self contained purge-air supply allows an area to be tested for up to 40 minutes without needing to 

take the monitor to an area of fresh air for zeroing purposes. The monitor self-zeros every 4 minutes to 
ensure accurate gas measurements. 
During normal operation the H-SEM unit displays the type of gas being monitored, along with displaying 

both the current gas level and the peak gas level detected in that area on its front panel LCD. The monitor 
retains a log of previous readings that can be easily accessed at a later time for analysis.  
Front panel indicators and an audible alarm are provided to signal gas alarm and instrument fault 
conditions. The audible alarm can temporarily be silenced by a push of a button, but will reactivate if the 

gas level is still above the alarm trip point after a user defined time period has elapsed. 
The H-SEM unit requires only minor periodic maintenance such as the occasional replacement of filters. 
The monitor incorporates active diagnostics that continuously check the system for proper operation. A 

front panel indicator is provided to alert an operator of system malfunctions, and fault codes are generated 

that enable the operator to identify the cause of the fault. 

1.9.2 Response to the Presence of Multiple Refrigerants 

The H-SEM is a refrigerant level monitor, not a gas analyzer. You must program the monitor to test for a 

specific refrigerant, and it will only return accurate concentration readings for that particular refrigerant. 

When other types of refrigerant gas are present, the monitor may return incorrect readings. 

Most applications only require detection of a single refrigerant and the problems that are associated with 

monitoring multiple gases are rarely an issue. If there is a possibility of multiple refrigerants leaking in 

the same sampling zone, then you should carefully consider which refrigerant compound you program the 
unit to monitor. 

Summary of Contents for H-SEM

Page 1: ...H SEM Halogen Safe Entry Monitor Instruction 3015 4584 Operation Maintenance Rev 0 November 2005 Product Leadership Training Service Reliability...

Page 2: ...ACCEPTED IN LIEU OF I ANY AND ALL OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND II ANY OBLIGATION...

Page 3: ...4 2 Charging the Battery 6 3 5 Display Screens 7 3 5 1 Initial Power Up 7 3 5 2 Filling the Purge Air Bag 7 3 5 3 Data Display Screen 7 3 5 4 Function Screens 7 3 6 Standby Mode 10 3 7 Working with t...

Page 4: ...tents H SEM ii Instruction 3015 4584 4 MAINTENANCE 15 4 1 Disassembly 15 4 2 Internal Particulate Hydrophobic Filter Replacement 16 5 PARTS AND SERVICE 17 5 1 Replacement Parts 17 5 2 Service Centers...

Page 5: ...d associated with the use of this equipment It calls attention to a procedure practice or condition or the like which if not correctly performed or adhered to could result in personal injury or death...

Page 6: ...ting due to changes in air pressure Deflate the bag by selecting the EMTYBAG function as described on Page 8 1 7 Unpacking and Initial Checks Your Halogen Safe Entry Monitor was carefully inspected at...

Page 7: ...e current gas level and the peak gas level detected in that area on its front panel LCD The monitor retains a log of previous readings that can be easily accessed at a later time for analysis Front pa...

Page 8: ...onds Temperature Drift 1 5 ppm per C between purge cycles System Noise Less than 40dB A at 10 ft 3 m Operating Temp 32 to 122 F 0 to 50 C Ambient Humidity 5 to 90 RH non condensing Altitude Limit 6 56...

Page 9: ...h air that is of the same temperature and relative humidity as the area being checked for gas DO NOT fill the purge air bag in an area that is contaminated with refrigerant gas After the purge air bag...

Page 10: ...le alarm will reactivate if the fault condition is not cleared within the time period set by the SILENCE function Page 9 Both the FAULT light and audible alarm will automatically turn OFF after the ca...

Page 11: ...ill automatically resume normal operation 3 5 3 Data Display Screen During normal operation the Data Display screen shows when the monitor is performing the following two functions MEASURE indicates t...

Page 12: ...g ON AUDALRM Allows the monitor s internal audible alarm to be associated with any function of the monitoring system Factory default is OFF Use the Keypad to select the desired audible alarm function...

Page 13: ...pe as specified on the sales order or a custom gas can be programmed into the monitor at a later time using the CUST K1 K2 K3 function Use the Keypad to select the desired gas type and then press ENTE...

Page 14: ...panel ALARM light will flash and the audible alarm if activated refer to AUDALRM function Page 8 will start beeping The ALARM light and audible alarm will automatically turn OFF once the detected gas...

Page 15: ...eme temperatures Use the DIAG function to check the Box Temperature 0002 Bench Temperature Fault Optical bench is outside normal operating range or IR detector has failed Check that the monitor is not...

Page 16: ...the First Diagnostic screen the user can toggle between displaying Bench Voltage Detector Temperature and Noise Box Temperature by pressing the Keypad Right button Bench Voltage This is the current pe...

Page 17: ...e Data Display screen press both the Keypad Left and Right buttons at the same time to display the first Function Menu screen Next use the Keypad buttons to move the arrow until it points to the P CHK...

Page 18: ...temperature drift IR source aging and pressure changes in order to keep the unit within factory accuracy specifications 3 13 1 Calibration Procedure The calibration factor is determined by sampling a...

Page 19: ...l chassis as follows Items Required Medium Phillips head screwdriver Procedure 1 Remove monitor and its battery pack from the soft carrying case 2 Unplug battery pack from monitor 3 Remove a total of...

Page 20: ...1 2 See illustration below and locate the internal particulate hydrophobic filter 3 Pull off tubing from both ends of filter and remove filter from instrument 4 Attach tubing to new filter being sure...

Page 21: ...rvice 5 1 Replacement Parts Item Description Part Number Halogen Safe Entry Monitor 3015 4484 External Filter 3015 3420 Internal Particulate Filter 07 1654 Battery Pack Charger 3015 4654 Battery Charg...

Page 22: ...il help bacharach inc com Canada Bacharach of Canada Inc 250 Shields Court Unit 3 Markham Ontario L3R 9W7 Canada Phone 905 470 8985 Fax 905 470 8963 Email bachcan idirect com M xico Bacharach de M xic...

Page 23: ...H SEM Parts and Service Instruction 3015 4584 19 Notes...

Page 24: ...1 Hunt Valley Circle New Kensington PA 15068 PH 724 334 5000 Toll Free 800 736 4666 FAX 724 334 5001 Website www bacharach inc com E mail help bacharach inc com Printed in U S A Registered Trademark o...

Reviews: