background image

Operation 

H-SEM 

10 

Instruction 3015-4584

 

3.6 Standby Mode 

After the monitor has been turned ON and allowed to warm up, the monitor can be placed into a Standby 

Mode, ready for operation without waiting an additional 15 minutes before using. While in standby, the 
monitor will stop taking gas samples and purging itself from the internal purge-air bag. 
Place the monitor into its Standby Mode by pressing both the 

ENTER

 and 

ESC

 buttons at the same time.  

 

 

Take the monitor out of standby by again pressing both the 

ENTER

 and 

ESC

 buttons.  

When the monitor is taken out of standby, the operator must refill the purge-air bag by pressing the 

ENTER

 button. Refilling the bag after coming out of standby re-establishes the monitor’s ambient 

pressure reading and ensures that the purge-air bag is filled at the start of taking gas measurements. 

WARNING:

 

Fill the purge-air bag in a clean, fresh-air environment that is at the same 

temperature and humidity as the area being checked for gas. DO NOT fill the bag in 
an area that is contaminated with refrigerant gas.

 

.

 

 

As soon as the purge-air bag has been refilled with fresh air, the 

Data Display

 screen appears and the 

monitor resumes normal operation. 

3.7  Working with the Gas Alarm 

3.7.1 Gas-Alarm Light 

If the detected gas level reaches the gas-alarm level as set by the 

ALARMLVL

 function (Page 8), the front 

panel 

ALARM

 light will flash and the audible alarm, if activated (refer to 

AUDALRM

 function Page 8), 

will start beeping. The 

ALARM

 light and audible alarm will automatically turn OFF once the detected gas 

level drops below the gas-alarm level. 

3.7.2 Silencing a Gas Alarm 

Pressing the 

SILENCE 

button while the alarm circuit is activated causes the internal audible alarm to 

turn OFF for a period of time as set by the 

SILENCE 

function (Page 9). The front panel 

ALARM

 light will 

continue to flash, however, as an indication that an alarm condition still exists. The audible alarm will 

reactivate at the end of the silence period if the detected gas level is still above the gas-alarm level. 

3.8  Working with System Faults 

3.8.1 Functional Overview 

If a system malfunction occurs (see Fault Code list, Page 11), the front panel 

FAULT

 light will flash and 

the audible alarm, if activated (refer to 

AUDALRM

 function Page 8), will start beeping. 

3.8.2 Clearing / Silencing a Fault Alarm 

The 

FAULT

 light and audible alarm will automatically turn OFF after the cause of the fault has been 

eliminated. 
Pressing the 

SILENCE 

button while a fault condition still exists causes the internal audible alarm to turn 

OFF for a period of time as set by the 

SILENCE 

function (Page 9). The front panel 

FAULT

 light will 

continue to flash, however, as a reminder that a fault condition still exists. The audible-alarm circuit will 

reactivate at the end of the silence period if the cause of the fault has not been corrected. 

    STAND BY 

PURGE AIR EMPTY 
<ENTER> TO FILL 

RECHARGING PURGE
AIR... 

Summary of Contents for H-SEM

Page 1: ...H SEM Halogen Safe Entry Monitor Instruction 3015 4584 Operation Maintenance Rev 0 November 2005 Product Leadership Training Service Reliability...

Page 2: ...ACCEPTED IN LIEU OF I ANY AND ALL OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND II ANY OBLIGATION...

Page 3: ...4 2 Charging the Battery 6 3 5 Display Screens 7 3 5 1 Initial Power Up 7 3 5 2 Filling the Purge Air Bag 7 3 5 3 Data Display Screen 7 3 5 4 Function Screens 7 3 6 Standby Mode 10 3 7 Working with t...

Page 4: ...tents H SEM ii Instruction 3015 4584 4 MAINTENANCE 15 4 1 Disassembly 15 4 2 Internal Particulate Hydrophobic Filter Replacement 16 5 PARTS AND SERVICE 17 5 1 Replacement Parts 17 5 2 Service Centers...

Page 5: ...d associated with the use of this equipment It calls attention to a procedure practice or condition or the like which if not correctly performed or adhered to could result in personal injury or death...

Page 6: ...ting due to changes in air pressure Deflate the bag by selecting the EMTYBAG function as described on Page 8 1 7 Unpacking and Initial Checks Your Halogen Safe Entry Monitor was carefully inspected at...

Page 7: ...e current gas level and the peak gas level detected in that area on its front panel LCD The monitor retains a log of previous readings that can be easily accessed at a later time for analysis Front pa...

Page 8: ...onds Temperature Drift 1 5 ppm per C between purge cycles System Noise Less than 40dB A at 10 ft 3 m Operating Temp 32 to 122 F 0 to 50 C Ambient Humidity 5 to 90 RH non condensing Altitude Limit 6 56...

Page 9: ...h air that is of the same temperature and relative humidity as the area being checked for gas DO NOT fill the purge air bag in an area that is contaminated with refrigerant gas After the purge air bag...

Page 10: ...le alarm will reactivate if the fault condition is not cleared within the time period set by the SILENCE function Page 9 Both the FAULT light and audible alarm will automatically turn OFF after the ca...

Page 11: ...ill automatically resume normal operation 3 5 3 Data Display Screen During normal operation the Data Display screen shows when the monitor is performing the following two functions MEASURE indicates t...

Page 12: ...g ON AUDALRM Allows the monitor s internal audible alarm to be associated with any function of the monitoring system Factory default is OFF Use the Keypad to select the desired audible alarm function...

Page 13: ...pe as specified on the sales order or a custom gas can be programmed into the monitor at a later time using the CUST K1 K2 K3 function Use the Keypad to select the desired gas type and then press ENTE...

Page 14: ...panel ALARM light will flash and the audible alarm if activated refer to AUDALRM function Page 8 will start beeping The ALARM light and audible alarm will automatically turn OFF once the detected gas...

Page 15: ...eme temperatures Use the DIAG function to check the Box Temperature 0002 Bench Temperature Fault Optical bench is outside normal operating range or IR detector has failed Check that the monitor is not...

Page 16: ...the First Diagnostic screen the user can toggle between displaying Bench Voltage Detector Temperature and Noise Box Temperature by pressing the Keypad Right button Bench Voltage This is the current pe...

Page 17: ...e Data Display screen press both the Keypad Left and Right buttons at the same time to display the first Function Menu screen Next use the Keypad buttons to move the arrow until it points to the P CHK...

Page 18: ...temperature drift IR source aging and pressure changes in order to keep the unit within factory accuracy specifications 3 13 1 Calibration Procedure The calibration factor is determined by sampling a...

Page 19: ...l chassis as follows Items Required Medium Phillips head screwdriver Procedure 1 Remove monitor and its battery pack from the soft carrying case 2 Unplug battery pack from monitor 3 Remove a total of...

Page 20: ...1 2 See illustration below and locate the internal particulate hydrophobic filter 3 Pull off tubing from both ends of filter and remove filter from instrument 4 Attach tubing to new filter being sure...

Page 21: ...rvice 5 1 Replacement Parts Item Description Part Number Halogen Safe Entry Monitor 3015 4484 External Filter 3015 3420 Internal Particulate Filter 07 1654 Battery Pack Charger 3015 4654 Battery Charg...

Page 22: ...il help bacharach inc com Canada Bacharach of Canada Inc 250 Shields Court Unit 3 Markham Ontario L3R 9W7 Canada Phone 905 470 8985 Fax 905 470 8963 Email bachcan idirect com M xico Bacharach de M xic...

Page 23: ...H SEM Parts and Service Instruction 3015 4584 19 Notes...

Page 24: ...1 Hunt Valley Circle New Kensington PA 15068 PH 724 334 5000 Toll Free 800 736 4666 FAX 724 334 5001 Website www bacharach inc com E mail help bacharach inc com Printed in U S A Registered Trademark o...

Reviews: